Manual

<6 OPERATION>
6-18
IM 01R01A11-01EN
(B) Flow (The uid velocity is sufcient)
1) Fill the owtube with uid. Be sure that the uid velocity is sufcient.
2
3
1
F060207.ai
Figure 6.2.7 “Flow/ No Flow” Screen
Table 6.2.6 “Flow/ No Flow” Screen
Item
No.
Item Name Description
1 Flow Selects theFlowtest condition.
2 Start Verication (*1)
Runs “Circuitcheck and “Device Status” check.
Those checks are performed automatically. The progress bar is displayed
at the bottom left.
3 Cancel Stops runningCircuit” check and “Device Status” check.
*1: When the AXF protect function is set to “Protect”, Figure 3.2.2 is displayed. Follow 3.2.2.
2) Select “Flow (1)” and click “Start Verication (2)”.
3) Before running “Circuit” check and “Device Status” check, Figure 6.2.8 is displayed auto-
matically.