Manual
<6 OPERATION>
6-4
IM 01R01A11-01EN
6.1.2 Save
By clicking the “Save” icon, the veri cation data is saved to the database.
After clicking “Save”, the screen changes as follows:
• When “No Flow + Flow (*1)” is not complete, see (A).
• When “No Flow + Flow” is complete, see (B).
*1: “No Flow + Flow”;
“Circuit” check and “Device Status” check for both “No Flow” and “Flow” conditions. Refer to 6.2.1.
F060105.ai
Figure 6.1.5 “Save” Icon
(A) When “No Flow + Flow” is not complete:
1) After clicking the “Save” icon, Figure 6.1.6 is displayed.
1 2 3
4
5
F060106.ai
Figure 6.1.6 “Saving Data When “No Flow + Flow” Uncompleted” Dialog
Table 6.1.3 “Saving Data When “No Flow + Flow” Uncompleted” Dialog
Item
No.
Item Name Description
1 Reason
Shows the reason or comment. It can be used as a marker when loading
veri cation data is selected. Refer to Figure 6.1.3.
Enter preferred text.
(Optional eld)
2 Saved as Installation Data.
Saves the data as “Installation Data”. See 7.1.
This eld is disabled until “No Flow + Flow” is complete.
3 Saved as Locked Data.
Saves the data as “Locked Data”. See 7.2.
To save the data as read-only, check the box.
(Always valid)
4 OK Saves the data.
5 Cancel Cancels saving the data and returns to “Main Screen”. See 6.1.