<1 QUICK START> 1-2
4.2
Leak-check all connections and ensure pressure ratings are not exceeded!
5.0
Power-Up
Make sure the power module door is closed. Do not open this door when the analyzer is powered. Apply
the AC power to the analyzer.
5.1 OpentheControlmoduledoor.InsidethismoduleusetheinternalOn-Offswitchtopower-uptheana-
lyzer(locatedlowerrighthandside).
5.2 ObservethevariousLEDclustersonthebackplaneandFPGAboards.
All blue LEDs located lower right side on the back-plane should be on.
5.3 ObservetheGreenpowerindicatorontheSBC.
5.4 ObservetheLEDsontheoptionalanalogI/Oboard.
6.0
Checking
Ifthereisaninstalledoptional6.5”DisplayandKeypad–ObservetheMainMenumessagesandstatus
information.
6.1 IfthereisnoinstalledUserInterface,thenconnectalaptopPCviaEthernettotheSBCmountedonthe
backplane.InitiatethesuppliedUltraVNCsoftwarefromthelaptoptoinitiateaVNCsessionwiththe
‘blind’ analyzer and observe the analyzer Main Menu via the laptop.
6.2
AT this time there may be alarm or warning messages due to low transmission, out of range
parameters or other – nal system conguration is still required!
6.3 IftheanalyzerdisplaysaWarning“ValidationRequired”thenthisindicatesthatthereisnotargetgas
absorption peak found at start-up. Either, shut down the analyzer, introduce some measured gas into the
owcellandre-start,orperformavalidation.Thiswillensurethattheanalyzeriscorrectlytunedtothe
measurement gas absorption peak. If this Warning cannot be cleared by either method, please contact
Yokogawa Laser Analysis Division or your local agent for further assistance.
7.0
Congure
BASIC
Bywayoftheappropriateuserinterface,thecorrectprocessparametersandotherparameterscannow
be entered.
7.1 EntertheBasicMenuandgotoCongure.
7.2
Gas Pressure
Enterinthecorrectprocessgaspressure(ifActive,seeAdvancedCongure).
7.3
Gas Temperature
Enterinthecorrectprocessgastemperature(ifActive,seeAdvancedCongure).
7.4 Ifanyotherparametersarerequiredtobeset(suchasanalogI/Oranges,alarmslevels,AutoValidation
sequences)thentheAdvancedMenuneedstobeaccessed.
Advanced Menu access is Password protected (default 1234, can be changed by user if neces-
sary) and should only be used by skilled and trained persons - Contact Yokogawa Laser Analysis
Division or Local Agent if any doubts!
GototheDatasectionunderBasicandconguretheappropriate‘RecordResultData’settings.Thiswill
ensuretheanalyzerstoresimportantinformationduringoperationthatmaybeusedtoverifyoperation/
status/diagnosticsandothertroubleshooting.
8.0
Congure
ADVANCED
Usingthecorrectpassword(Default1234),enterintotheAdvancedMenu,thentheCongure.
8.1 Select the desired measurement units for path length, pressure and temperature.