Owner's manual
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
<1 QUICK START> 1-2
4.2
Leak-check all connections and ensure pressure ratings are not exceeded!
5.0
Power-Up
Make sure the power module door is closed. Do not open this door when the analyzer is powered. Apply
the AC power to the analyzer.
5.1 OpentheControlmoduledoor.InsidethismoduleusetheinternalOn-Offswitchtopower-uptheana-
lyzer(locatedlowerrighthandside).
5.2 ObservethevariousLEDclustersonthebackplaneandFPGAboards.
All blue LEDs located lower right side on the back-plane should be on.
5.3 ObservetheGreenpowerindicatorontheSBC.
5.4 ObservetheLEDsontheoptionalanalogI/Oboard.
6.0
Checking
Ifthereisaninstalledoptional6.5”DisplayandKeypad–ObservetheMainMenumessagesandstatus
information.
6.1 IfthereisnoinstalledUserInterface,thenconnectalaptopPCviaEthernettotheSBCmountedonthe
backplane.InitiatethesuppliedUltraVNCsoftwarefromthelaptoptoinitiateaVNCsessionwiththe
‘blind’ analyzer and observe the analyzer Main Menu via the laptop.
6.2
AT this time there may be alarm or warning messages due to low transmission, out of range
parameters or other – nal system conguration is still required!
6.3 IftheanalyzerdisplaysaWarning“ValidationRequired”thenthisindicatesthatthereisnotargetgas
absorption peak found at start-up. Either, shut down the analyzer, introduce some measured gas into the
owcellandre-start,orperformavalidation.Thiswillensurethattheanalyzeriscorrectlytunedtothe
measurement gas absorption peak. If this Warning cannot be cleared by either method, please contact
Yokogawa Laser Analysis Division or your local agent for further assistance.
7.0
Congure
BASIC
Bywayoftheappropriateuserinterface,thecorrectprocessparametersandotherparameterscannow
be entered.
7.1 EntertheBasicMenuandgotoCongure.
7.2
Gas Pressure
Enterinthecorrectprocessgaspressure(ifActive,seeAdvancedCongure).
7.3
Gas Temperature
Enterinthecorrectprocessgastemperature(ifActive,seeAdvancedCongure).
7.4 Ifanyotherparametersarerequiredtobeset(suchasanalogI/Oranges,alarmslevels,AutoValidation
sequences)thentheAdvancedMenuneedstobeaccessed.
Advanced Menu access is Password protected (default 1234, can be changed by user if neces-
sary) and should only be used by skilled and trained persons - Contact Yokogawa Laser Analysis
Division or Local Agent if any doubts!
GototheDatasectionunderBasicandconguretheappropriate‘RecordResultData’settings.Thiswill
ensuretheanalyzerstoresimportantinformationduringoperationthatmaybeusedtoverifyoperation/
status/diagnosticsandothertroubleshooting.
8.0
Congure
ADVANCED
Usingthecorrectpassword(Default1234),enterintotheAdvancedMenu,thentheCongure.
8.1 Select the desired measurement units for path length, pressure and temperature.