Owner's manual
<9 DATA FILES AND FORMAT> 9-4
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
system .bak
System History Back-Up
A back-up of historical log of
congurationchanges
SystemCongurationHistory-Whenthesystem.hisleexceeds
100KBsizethecontentsissavedtothis.bakleandthe.hisle
isemptied.This.bakleisintheformofASCIIdatalesthatcan
beopenedwithMicrosoft“Notepad”assimple.txtleformats.
The content can then be copied and pasted into Microsoft Excel
spread.Eachcongurationchangeisloggedwiththedate(MM-
DD-YYYY)andtimethatitoccurred,theparameter(e.g.opl)and
the new value. Also, whenever the analyzer is started-up or shut-
downthe‘TruePeakOpen/Close”parameterwillbelogged.
Example:
11-20-200608:13:10opl(inch)39.9643.31
11-20-200608:13:26temperature(F)509.086.0
11-20-200608:13:40pressure(psi)14.5014.50
system .usr
System User
Systemcongurationasshippedfrom
the factory
Thiscontainsall‘Factory’congurationsettingsastheanalyzer
was shipped. This is used for reference purposes only.
valspe .his
Validation Spectra
AllspectracapturedduringOn-Lineand
Off-LineValidations
SpectracapturedforhistoricalValidations(On-LineandOff-Line)
arestoredinthisle.
valspe .bak
Validation Spectra Back-Up
Backed-upspectracapturedduringOn-
LineandOff-LineValidations
Whenthevalspe.hisleexceeds1MBsizethecontentsis
savedtothis.bakleandthe.hisleisemptied.This.bakleis
SpectracapturedforhistoricalValidations(On-LineandOff-Line)
arestoredinthisle.
zero00 .spe
Zero Calibration Spectra
AbsorptionSpectrumandcoefcientsat
time of calibration
Thisleisessentialtotheanalyzercalibration-ifthisledoes
notexistoriscorrupted,modiedorotherwisetamperedwith
then the analyzer calibration is invalid. It contains essential
informationrelatingtotheZeroCalibrationoftheanalyzer.
zero01
to
zero10
..spe
Historical Zero Calibration
Spectra
PreviousAbsorptionSpectrum
andcoefcientsattimeofprevious
calibrations
zero01throughzero09arepreviousleswithzero01beingthe
mostrecent.Themostrecentpreviouszero01canberestored
intheanalyzeras'PreviousCalibration'.Thefactorycalibration
whichcanalsoberestoredintheanalyzerisnamedzero10.
zeroref .cfg
System Conguration
AtthetimeofcurrentZeroCalibration
SystemCongurationatthetimeofcurrentZeroCalibration
(zero00.spe)
memory .res
File Size ManagementforResults
FACTORYPERSONNELONLY-usedbyTruePeaksoftwareto
managethe.resles
memory .spe
File Size Management for Spectra
FACTORYPERSONNELONLY-usedbyTruePeaksoftwareto
managethe.speles