Owner's manual

<9 DATA FILES AND FORMAT> 9-3
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
calibr .bak
Calibration History Back-Up
A back-up of historical log of analyzer
calibrations
CalibrationHistory-Whenthecalibr.hisleexceeds100KBsize
thecontentsissavedtothis.bakleandthe.hisleisemptied.
This.bakleisintheformofASCIIdatalesthatcanbeopened
withMicrosoft“Notepad”assimple.txtleformats.Thecontent
can then be copied and pasted into Microsoft Excel spread.
EachCalibrationand/orValidationeventisloggedwiththedate
(MM-DD-YYYY)andtimethatitoccurred,theMode(Calibration
orValidation),thespecictype(TransmissionCalibration,Zero
Calibration,etc.)andthe‘K’(constant)factorusedfortheevent
Example:
11-14-200714:11:19span_calibrate612814.899250.84
152665.88
11-14-200714:50:16transmission_cal14.95
calibr .pik
Calibration Pick List - Factory
Use Only
FACTORYPERSONNELONLY
S .dat
Chemometric Model File
Parentspectraforchemometricmodelusedonlyforspecialized
applicationsthatutilizetheTruePeakCLSmeasurement
capability.
span00 .spe
Span Calibration Spectra
AbsorptionSpectrumandcoefcients
at time of calibration
Thisleisessentialtotheanalyzercalibration-ifthisledoes
notexistoriscorrupted,modiedorotherwisetamperedwith
then the analyzer calibration is invalid. It contains essential
information relating to the Span Calibration of the analyzer.
span01
to
span10
.spe
Historical Span Calibration
Spectra
PreviousAbsorptionSpectrum
andcoefcientsattimeofprevious
calibrations
span01throughspan09arepreviousleswithspan01beingthe
mostrecent.Themostrecentpreviousspan01canberestored
intheanalyzeras‘PreviousCalibration’.Thefactorycalibration
whichcanalsoberestoredintheanalyzerisnamedspan10.
system .cfg
System Conguration
Allessentialanalyzer&installation
specicparametersrequiredbythe
analyzerforthegivenapplication&
installation
Thisleisessentialtotheanalyzer-ifthisledoesnotexist
oriscorrupted,modiedorotherwisetamperedwiththen
theanalyzercannotfunction.Itcontainsanalyzerspecic
parameters relating to every detail of the measurement,
calibrations,compensationfactors,I/Ocongurations,Valve
controlcongurations,signalprocessing,etc.Thisleisusedby
FACTORYPERSONNELONLYtoevaluateifthecongurationis
appropriate for the analyzers intended use.
system .his
System History
Ahistoricallogofconguration
changes
SystemCongurationHistory-intheformofASCIIdatales
thatcanbeopenedwithMicrosoft“Notepad”assimple.txt
leformats.Thecontentcanthenbecopiedandpastedinto
MicrosoftExcelspread.Eachcongurationchangeislogged
withthedate(MM-DD-YYYY)andtimethatitoccurred,the
parameter(e.g.opl)andthenewvalue.Also,wheneverthe
analyzerisstarted-uporshut-downthe‘TruePeakOpen/Close”
parameter will be logged.
Example:
11-20-200608:13:10opl(inch)39.9643.31
11-20-200608:13:26temperature(F)509.086.0
11-20-200608:13:40pressure(psi)14.5014.50