Owner's manual
<9 DATA FILES AND FORMAT> 9-3
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
calibr .bak
Calibration History Back-Up
A back-up of historical log of analyzer
calibrations
CalibrationHistory-Whenthecalibr.hisleexceeds100KBsize
thecontentsissavedtothis.bakleandthe.hisleisemptied.
This.bakleisintheformofASCIIdatalesthatcanbeopened
withMicrosoft“Notepad”assimple.txtleformats.Thecontent
can then be copied and pasted into Microsoft Excel spread.
EachCalibrationand/orValidationeventisloggedwiththedate
(MM-DD-YYYY)andtimethatitoccurred,theMode(Calibration
orValidation),thespecictype(TransmissionCalibration,Zero
Calibration,etc.)andthe‘K’(constant)factorusedfortheevent
Example:
11-14-200714:11:19span_calibrate612814.899250.84
152665.88
11-14-200714:50:16transmission_cal14.95
calibr .pik
Calibration Pick List - Factory
Use Only
FACTORYPERSONNELONLY
S .dat
Chemometric Model File
Parentspectraforchemometricmodelusedonlyforspecialized
applicationsthatutilizetheTruePeakCLSmeasurement
capability.
span00 .spe
Span Calibration Spectra
AbsorptionSpectrumandcoefcients
at time of calibration
Thisleisessentialtotheanalyzercalibration-ifthisledoes
notexistoriscorrupted,modiedorotherwisetamperedwith
then the analyzer calibration is invalid. It contains essential
information relating to the Span Calibration of the analyzer.
span01
to
span10
.spe
Historical Span Calibration
Spectra
PreviousAbsorptionSpectrum
andcoefcientsattimeofprevious
calibrations
span01throughspan09arepreviousleswithspan01beingthe
mostrecent.Themostrecentpreviousspan01canberestored
intheanalyzeras‘PreviousCalibration’.Thefactorycalibration
whichcanalsoberestoredintheanalyzerisnamedspan10.
system .cfg
System Conguration
Allessentialanalyzer&installation
specicparametersrequiredbythe
analyzerforthegivenapplication&
installation
Thisleisessentialtotheanalyzer-ifthisledoesnotexist
oriscorrupted,modiedorotherwisetamperedwiththen
theanalyzercannotfunction.Itcontainsanalyzerspecic
parameters relating to every detail of the measurement,
calibrations,compensationfactors,I/Ocongurations,Valve
controlcongurations,signalprocessing,etc.Thisleisusedby
FACTORYPERSONNELONLYtoevaluateifthecongurationis
appropriate for the analyzers intended use.
system .his
System History
Ahistoricallogofconguration
changes
SystemCongurationHistory-intheformofASCIIdatales
thatcanbeopenedwithMicrosoft“Notepad”assimple.txt
leformats.Thecontentcanthenbecopiedandpastedinto
MicrosoftExcelspread.Eachcongurationchangeislogged
withthedate(MM-DD-YYYY)andtimethatitoccurred,the
parameter(e.g.opl)andthenewvalue.Also,wheneverthe
analyzerisstarted-uporshut-downthe‘TruePeakOpen/Close”
parameter will be logged.
Example:
11-20-200608:13:10opl(inch)39.9643.31
11-20-200608:13:26temperature(F)509.086.0
11-20-200608:13:40pressure(psi)14.5014.50