Followingtheonscreendirections,enterinthe:
Pressure=CellPressure
Temperature=CellTemperature
Celllength=40”
Concentration of the target gas inside the gas
owcell.PressEntertoproceed.
Waitforthereadingtostabilize.Press9to
proceed.
AUTOMATICOFFLINEVALIDATIONS
TheOfineValidationscanalsobe
automaticallycongured.Refertosection
5.5.3fordetails.
Local Initiate will start the automatic online
validation sequence when selected. It will use
theexisting‘Settings’(seebelowfordetailson
‘Settings’).
RemoteInitiatewillenable/disablemonitoring
oftheselected“RemoteInitiate”contactson
TB-2withintheLaunchUnit.Whenenabled,
the analyzer will detect the chosen contact
closure and automatically start the online
validation sequence.
TimeInitiatewillallowinputofaspecied
time to automatically start online validation
sequenceonceeveryday,everyweek,every2
weeks,orevery4weeks.