Owner's manual
<7 ROUTINE MAINTENANCE> 7-12
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
Followingtheonscreendirections,enterinthe:
Pressure=CellPressure
Temperature=CellTemperature
Celllength=40”
Concentration of the target gas inside the gas
owcell.PressEntertoproceed.
Waitforthereadingtostabilize.Press9to
proceed.
AUTOMATICOFFLINEVALIDATIONS
TheOfineValidationscanalsobe
automaticallycongured.Refertosection
5.5.3fordetails.
Local Initiate will start the automatic online
validation sequence when selected. It will use
theexisting‘Settings’(seebelowfordetailson
‘Settings’).
RemoteInitiatewillenable/disablemonitoring
oftheselected“RemoteInitiate”contactson
TB-2withintheLaunchUnit.Whenenabled,
the analyzer will detect the chosen contact
closure and automatically start the online
validation sequence.
TimeInitiatewillallowinputofaspecied
time to automatically start online validation
sequenceonceeveryday,everyweek,every2
weeks,orevery4weeks.