Owner's manual
<6 BASIC OPERATION> 6-22
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
6.5.6 OPTIONAL Virtual Analyzer Controller (VAC) Operating Software Guide
OnceanEthernetconnectionhasbeenestablished,the
VACsoftwarecanbestarted.Toloadthesoftwareonto
alaptopordesktopPC,followthesesteps:
Create folder under c:\asiini
CopyalllesfromtheCDROMtothisfolder
Create a shortcut from the desktop by browsing for the
ASI-VAC.exele
TesttheshortcutbydoubleclickingonitandtheVAC
software should open
Oncethesoftwareisopen,followtheon-screenoptions.
Functionkeyshavebeenpre-assignedtocertaintasks
allowingaRIUkeypadtoperformallnecessarytasks.
The primary functions of the software are:
Create a virtual network computing connection to an
analyzer thus allowing for control of the analyzer, typically
for start-up, service, calibration, etc.
AllowforletransferfromananalyzertoalocalUSBport
(formemorydevice)
Createconnectionsbynameand/orIPaddress
OFFLINECALIBRATIONS–useforallTDLS220
ZEROCALIBRATION–manualorautomaticcalibrationof
Zero.
ZEROOFFSET–allowsmanualadjustmentofZeroby
applying a concentration offset.
SPANCALIBRATION–manualorautomaticcalibrationof
Span.
TRANSMISSION–adjustmentoftransmissionvalue.
DARKCURRENT–calibrationofdetectordarkcurrent.
PEAKSEARCH–initiatesasystemscanofabsorption
peaks to validate current peak selection is correct.
Afterselecting“CopyData–F3”youwillbeallowedto
selecttheanalyzer(withdescriptionandIPaddress)you
wish receive data from.
This will initiate a data transfer for all results and
congurationlesstoredontheanalyzer.