Owner's manual

<6 BASIC OPERATION> 6-12
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
ADVANCED CALIBRATE & VALIDATE MENU
OFFLINE CALIBRATIONS–allowszerocalibration,
zero offset, span calibration, transmission adjustment,
dark current calibration, and peak search.
OFFLINE VALIDATIONS–allowsmanualorautomatic
congurationofcheckgases1through3.
ONLINE VALIDATIONS – DO NOT USE THIS
FEATURE on TDLS220
OFFLINE CALIBRATIONS–useforallTDLS220
ZERO CALIBRATION–manualorautomatic
calibrationofZero.
ZERO OFFSET–allowsmanualadjustmentofZeroby
applying a concentration offset.
SPAN CALIBRATION–manualorautomatic
calibration of Span.
TRANSMISSION–adjustmentoftransmissionvalue.
DARK CURRENT–calibrationofdetectordark
current.
PEAK SEARCH–initiatesasystemscanofabsorption
peaks to validate current peak selection is correct.
OFFLINE VALIDATIONS
CHECK GAS 1 to 3 – allows manual or automatic
congurationupto3checkgasses.
Seesection5.5.4.