Owner's manual
<6 BASIC OPERATION> 6-11
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
VALVE CONTROL
Typicallyusedwhentheanalyzerisconguredwitha
owcellinanofineapplication.
VALVE CONTROL
The valve control includes manual and automatic
control.
MANUAL–allowsformanualON/OFFcontrolofthe
valve driver output signals.
TIME SEQUENCE–allowsautomaticvalveswitching
based on time sequence. It is normally used for
measurement that needs stream switching.
REMOTE OVERRIDE–allowsremotevalvecontrolby
digital contact.
LASER SPECTRA & CONTROL
DisplaysRawDetectorSignalandAbsorption
SpectrumaswellasGasConcentration,Gas
Temperature,GasPressure,Transmission, Laser
Temperature and Peak Center Position.
CAPTURE–allowsamanualspectracapture(userwill
bepromptedtoenterauniquelenameforcaptured
spectra).
ForDisplayOnly:
CONTROL MODE–allowsselectionofAutomatic
(lasertemperatureiscontrolledtokeeppeakcentered
usingpeakcenterpositionassetpoint)orManual
(lasertemperatureiscontrolledusingintegrallaser
temperaturesensor)–ONLYADJUSTWITHFACTORY
ASSISTANCE.
LASER TEMP–Inmanualmodeallowsadjustmentof
lasertemperature–ONLYADJUSTWITHFACTORY
ASSISTANCE.
LTSP LIMITS –settingofguardlimitsforlaser
temperaturesetpoint–ONLYADJUSTWITH
FACTORYASSISTANCE.
CUR CENTER–settingofcenterpointforlaser
currentramp–ONLYADJUSTWITHFACTORY
ASSISTANCE.