<6 BASIC OPERATION> 6-10
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
CONCENTRATION OUT OF RANGE –upperand
lower levels at which process gas concentration read-
ing will trigger a warning.
BOARD TEMPERATURE OUT OF RANGE–warning
that analyzer internal temperature is excessive.
VALIDATION FAILURE–avalidationfailurewilltrigger
awarning;therearenosettingsassociatedwiththis.
CHANNEL 2 FAULTS menu allows setting of various
analyzerFAULTconditions.FAULTisaneventthatwill
eliminate the measurement integrity, it is an indication
that maintenance is required and the analyzer is not
operational.
– ONLY ADJUST WITH FACTORY ASSISTANCE –
FAULT CONDITIONS ARE CRITICAL SETTINGS
THAT CAN RESULT IN DAMAGE TO THE ANALYZ-
ER IF IMPROPERLY PROGRAMMED.
LASER TEMPERATURE OUT OF RANGE–upper
and lower fault conditions for laser temperature.
DETECTOR SIGNAL HIGH–upperrawdetector
signal limit.
DETECTOR SIGNAL LOST–lowerrawdetector
signal limit.
OUTLIER REJECTION –thresholdofspectrumnoise
that will result in rejection of measurement.
PEAK CENTER OUT OF RANGE–lossofpeakcenter
control.
SYSTEM
Some settings are not adjustable by user, user
adjustment is possible for:
PASSWORD–changespasswordforADVANCED
menu access.
DATE & TIME–changesanalyzerdateandtime.