Owner's manual
<6 BASIC OPERATION> 6-10
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
CONCENTRATION OUT OF RANGE –upperand
lower levels at which process gas concentration read-
ing will trigger a warning.
BOARD TEMPERATURE OUT OF RANGE–warning
that analyzer internal temperature is excessive.
VALIDATION FAILURE–avalidationfailurewilltrigger
awarning;therearenosettingsassociatedwiththis.
CHANNEL 2 FAULTS menu allows setting of various
analyzerFAULTconditions.FAULTisaneventthatwill
eliminate the measurement integrity, it is an indication
that maintenance is required and the analyzer is not
operational.
– ONLY ADJUST WITH FACTORY ASSISTANCE –
FAULT CONDITIONS ARE CRITICAL SETTINGS
THAT CAN RESULT IN DAMAGE TO THE ANALYZ-
ER IF IMPROPERLY PROGRAMMED.
LASER TEMPERATURE OUT OF RANGE–upper
and lower fault conditions for laser temperature.
DETECTOR SIGNAL HIGH–upperrawdetector
signal limit.
DETECTOR SIGNAL LOST–lowerrawdetector
signal limit.
OUTLIER REJECTION –thresholdofspectrumnoise
that will result in rejection of measurement.
PEAK CENTER OUT OF RANGE–lossofpeakcenter
control.
SYSTEM
Some settings are not adjustable by user, user
adjustment is possible for:
PASSWORD–changespasswordforADVANCED
menu access.
DATE & TIME–changesanalyzerdateandtime.