Owner's manual
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
SAFETYshouldbeconsideredrstandforemostimportancewhenworkingontheequipmentdescribedinthismanual.Allpersonsusingthis
manual in conjunction with the equipment must evaluate all aspects of the task for potential risks, hazards and dangerous situations that may
existorpotentiallyexist.PleasetakeappropriateactiontopreventALLPOTENTIALACCIDENTS.
1) Safe lifting and carrying
If it is necessary to relocate the analyzer, it should be lifted by at least two people wearing protective gloves and steel toe boots.
The analyzer should be always transported either in a vertical position as shown in the picture, or in a horizontal position with the mounting plate
facing down. If transported horizontally, it must be held by the mounting plate only. In vertical position, it is acceptable to hold the analyzer by the
elementsinthe“grip”area.Neverholdtheanalyzerbyapplyingforcetotheelementsinthe“Donotgrip”zoneortotheconectingcablesand
tubes. If a winch or another hoisting device is used, it can be hooked to the upper mounting holes.
2) Electrical hazard
The areas of potentially hazardous voltage are labeled with this sign:
Theanalyzerispoweredbyeithera~220V,50Hzor~120V,60Hz(modelspecic).Duringnormaloperationthisvoltageisnotaccessiblefrom
outsideoftheenclosure.However,enduserisresponsibleforconnectingthemainstotheanalyzerterminalsinthePower/TemperatureController
box(rightbottomblockinthediagramabove).
THEWIRESBEINGCONNECTEDMUSTBEDISCONNECTEDFROMANYVOLTAGESOURCESDURINGTHISPROCEDURE.
PROTECTIVEGROUNDWIREMUSTBECONNECTEDTOTHEDESIGNATEDTERMINALLABELEDBYTHECORRESPONDINGSYMBOL.
DONOTOPENTHEPOWERMODULEENCLOSUREWHENTHEANALYZERISENERGIZED.
SomemaintenanceandtroubleshootingoperationsrequireopeningtheElectronicsControllerbox(leftbottomblockinthediagramabove)while
theanalyzerispowered.Thehighestvoltagepresentinthisboxis24Vandisnothazardous.Ahighervoltage(+65V,10mA)
is shielded with an isolating plate and labeled. .
3) Thermal hazard
Thegascellcanbeheatedupto+120ºCandcausethermalinjurytounprotectedskinatdirectcontact.Innormalcongurationthecellis
wrappedinathermalisolationjacketthatpreventssuchincidents.DONOTREMOVETHERMALJACKETWHENTHEANALYZERISPOWERED.
Priortoperformingserviceoperationsthatrequireremovalofthethermaljacket,powertheanalyzerdownandwait30minutestoletthe
temperature drop to a safe level.
iii