Owner's manual
<6 BASIC OPERATION> 6-9
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
SYSTEM I/O – ANALOG INPUT
Pre-calibratedatfactoryandnotnormallyrequired.
Allowscalibrationof4to20mAinputchannels;follow
onscreen instruction.
SYSTEMI/O-DIGITALOUTPUT
SettingofDigitalOutputassignments(DO1-3)
CHANNEL1WARNINGS–Settingoflevelsthatwill
triggeranalyzerwarningandsubsequentDO.
CHANNEL2FAULTS–Settingoflevelsthatwilltrigger
analyzerfaultandsubsequentDO.
CHANNEL3USERALARM–SettingofeitherCon-
centrationorTransmissionlevel(highorlow)thatwill
triggeranalyzerwarningandsubsequentDO.
FIELDLOOPCHECK–allowswiringconnectioncheck
duringeldinstallation.
CHANNEL 1 WARNINGS menu allows setting of
variousanalyzerWARNINGconditions.Warning
is an event that will reduce but not eliminate the
measurement integrity, it is an indication that
maintenance is required but the analyzer is still
operational.
DETECTOR SIGNAL LOW–levelatwhichlow
detectorsignalwilltriggerawarning–ONLYADJUST
WITHFACTORYASSISTANCE.
TRANSMISSION LOW–levelatwhichlow
transmissionwilltriggerawarning–ONLYADJUST
WITHFACTORYASSISTANCE.
SPECTRUM NOISE HIGH–levelatwhichexcessive
spectrumnoisewilltriggerawarning-–ONLY
ADJUSTWITHFACTORYASSISTANCE.
PROCESS PRESSURE OUT OF RANGE –upperand
lower levels at which process gas pressure reading will
trigger a warning.
PROCESS TEMPERATURE OUT OF RANGE–upper
and lower levels at which process gas temperature
reading will trigger a warning.