Owner's manual

<6 BASIC OPERATION> 6-8
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
UNITS–selectionofunitsforpathlength(in,ft,cm,m),
pressure(psiA,barA,kPa,torr,atm)andtemperature(ºF,
ºC,ºK).
SYSTEM I/O–allowssetupandassigningofanalyzer
AnalogandDigitalI/O.
SYSTEM–displaysanalyzerinformation(serialnumber,
Fatdate,password,softwareversion,launch/detectunit
temperatures),allowssettingofdate/time,gastype.
VALVE CONTROL¬–allowsformanualand/orautomatic
control of the valve driver output signals.
SIGNAL PROCESSING –Factorysetparametersonly
LASER SPECTRA & CONTROL–displaysspectraand
allows manual control of laser.
SYSTEM I/O - ANALOG OUTPUT
CHANNEL 1 to 3 –conguringeach4to20mAchannel
tooutputConcentration,Transmission,GasTemperature,
GasPressureorNone.
WARNING MODE–settingofmAoutputresponseduring
analyzerwarnings(Block,Track,Hold).
FAULT MODE–settingofmAoutputresponseduring
analyzerwarnings(Block,Track,Hold).
BLOCK MODE LEVEL–speciestheanalogoutputlevel
inblockmode(high20mAorlow3.3mA).
FIELD LOOP CHECK–allowsspecied4-20mAoutput
levels to check and distinguish between the three analog
outputconnections;selectanalogoutputchannel1,2,or
3tocheckandinputnewvaluetooutputandpressenter
to activate.
AO CH CALIBRATION–Pre-Calibratedatfactoryandnot
normallyrequired.Allowscalibrationof4to20mAoutput
channels;followonscreeninstructions.