Owner's manual
<6 BASIC OPERATION> 6-7
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
The TREND SCREENisidenticalforBASICor
ADVANCEDMENUS.Itallowstheusertotrenduptothe
last750minutes(ofcurrentday)ofdatafor:
REFRESH - The trend will not update automatically, use
the refresh button to update the trend
GAS 1–analyzerreadingofgas1concentration
STDEV1–thestandarddeviationof25consecutive
concentrationreadingsofGas1
GAS 2–analyzerreadingofgas2concentration
STDEV2–thestandarddeviationof25consecutive
concentrationreadingsofGas2
TRANS.–transmission%oflaserlightthroughthe
process
LTS–analyzerlasertemperaturesetpoint
LT–analyzerlasertemperature
PCP–peakcenterpositionfortheabsorptionpeak
TEMP–processgastemperature
PRES–processgaspressure
Alongside the selection buttons the current value is
displayed. When selecting the information to trend user
will be prompted to enter minimum value, maximum value
and time to trend.
ADVANCED CONFIGURE MENU
PROCESS PATH LENGTH–DONOTADJUST,factory
settypical40”(distancelaserbeamisexposedto
processgas).
PROCESS PRESSURE–allowsselectionofFIXED(value
enteredintosoftware),ACTIVE(analyzerfedpressure
valuefromexternaltransducer),orCONTROL(cell
pressure controlled at a desired value with a proportional
valveandpump).InActivemode,aBack-Upvaluecanbe
entered, in case of active input failure.
TEMPERATURE–allowsselectionofFIXED(value
enteredintosoftware),ACTIVEINPUT(analyzerfed
temperaturevaluefromexternaltransducer),ACTIVE
AMBIENT(ambientgastemperaturederivedfrominternal
sensor),ACTIVEPEAKS(valuecalculatedfromthe
measurementspectrum),orCONTROL(celltemperature
controlledatadesiredvaluewithaheaterandrelay).In
ActiveInputmode,aBack-Upvaluecanbeentered,in
case of active input failure.
NON-PROCESS PARAMETER–iftheanalyzerisnot
purged with a gas that is free of the component being
measured(i.e.oxygen),thisallowsusertoenterthePath
Length,Pressure,TemperatureandConcentrationof
the purge gas that is present in the non-measurement
opticalpath(internalanalyzeropticalpath)..Fordetails,
seesection4.4.UsethisfeatureiftheTDLS220isbeing
purgedwithInstrumentAir(I/A)