Owner's manual

<6 BASIC OPERATION> 6-5
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
6.2 Software Guide
MAIN MENU
DisplayofConcentration&Transmission
StatusWindow–noticationofinitiating,working
properly, warnings or faults
GasTemperature
GasPressure
SelectionofBasicorAdvancedMenu
ActiveAlarmDisplayButton
AnalyzerShutDownButton
Tag number and serial number
Analyzer date and time
AfterselectionofeitherBasicorAdvancedMenu
youwillseetheOutputSelectionscreen.
This allows control of the analog output while the user
is working in the analyzer software.
• Blockwillholdoutputsat3.8mAuntil
return to Main Screen
• Trackwillallowoutputstocontinueto
report concentration and transmission
until return to Main Screen
• Holdwillholdoutputsattheircurrent
value until return to Main Screen
BASIC MENU
Congure–allowssettingofPathLength,GasTem-
perature,GasPressure
ViewSpectra–userwillselectdisplayofrawdetector
signal or absorption spectra
Data–AlarmHistory,CalibrationHistory,Record
ResultData
Trends–allowsforthedisplayingofdatainatrend
format