Owner's manual
<6 BASIC OPERATION> 6-5
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
6.2 Software Guide
MAIN MENU
DisplayofConcentration&Transmission
StatusWindow–noticationofinitiating,working
properly, warnings or faults
GasTemperature
GasPressure
SelectionofBasicorAdvancedMenu
ActiveAlarmDisplayButton
AnalyzerShutDownButton
Tag number and serial number
Analyzer date and time
AfterselectionofeitherBasicorAdvancedMenu
youwillseetheOutputSelectionscreen.
This allows control of the analog output while the user
is working in the analyzer software.
• Blockwillholdoutputsat3.8mAuntil
return to Main Screen
• Trackwillallowoutputstocontinueto
report concentration and transmission
until return to Main Screen
• Holdwillholdoutputsattheircurrent
value until return to Main Screen
BASIC MENU
Congure–allowssettingofPathLength,GasTem-
perature,GasPressure
ViewSpectra–userwillselectdisplayofrawdetector
signal or absorption spectra
Data–AlarmHistory,CalibrationHistory,Record
ResultData
Trends–allowsforthedisplayingofdatainatrend
format