Owner's manual

<6 BASIC OPERATION> 6-4
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
Optional “Mini-Display 4 line x 20 character” Software Map
Display Text Description
Line1-Measurement O2xx.x% Measured gas and unit of measurement
Line2–TransmissionorSec-
ondGasMeasurement
Transmissionxx.x% Laser light transmission strength
(0-100%range)
CH4xx.x% Second measurement gas and unit
Line3-Status
Initializing…… shown during the power-up and initialization of the
analyzer
SystemOK NormalOperationconditionwithnoactivealarms
WARNINGDetSigLow
WARNINGConditions
WARNINGTransLow
WARNINGSpectrNoise
WARNINGGasPres
WARNINGGasTemp
WARNINGGasLevel
WARNINGBoardTemp
WARNINGValFailure
WARNINGValRequire
FAULTLaserTemp
FAULTConditions
FAULTDetSigHigh
FAULTDetSigLost
FAULTOutlier
FAULTPeakCenter
ZeroCalibrating…
ValidationStatus
Span Calibrating…
OfineValidating…
OnlineValidating…
Data Transferring…
Data Transfer Status
Transfer Success
TransferFailure
Line4-Information
YOKOGAWATDLS220 AnalyzerName
SN76-1xxx-05-xx AnalyzerSerialNo.
AO1:CONCxx-xx%/ppm/ppb
Congured4-20mAoutputfor
AO1,AO2,&AO3
AO2:TRANSxx-xx%
AO3:TEMPxx-xxF/C/K
10.0.0.35 StaticIPAddress
TEMPAct/Con/FixxxF/C/K ProcessGasTemperatureusedforgasconcentration
calculation
PRESAct/Con/Fixxx.xPsiA/
BarA
ProcessGasPressureusedforgasconcentration
calculation
OPLxx.xin/cm OpticalPathLengthoverwhichtheanalyzerismeasuring
the target gas
Launch xx C Launch unit internal temperature
Detect xx C Detect unit internal temperature