Owner's manual

<5 INSTALLATION AND WIRING> 5-1
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
5 Installation and Wiring
5.1 Mounting of the Analyzer
RefertotheSAFETYsection(Safehandlingandrelocation)forinstructionsonholdingandmovingthe
analyzer.
TheTDLS220issuitableforwallmountingbymeansofthefourcornerlocatedmountingholesasshown
below:Theholedare½”sizedfortypically3/8”or¼”boltsandshouldbeusedwithappropriatesized
atandlockingwashers.Ensuretheanalyzerissecuringfastenedandthatthereissuitableaccessfor
maintenance, etc.
Theanalyzerisdesignedforoperationoutsidebuildings,atnormalenvironmentalconditions(see
Environmentalspecications,page11).However,itmustbeprotectedfromdirectrainandsnowfall.
Mounting in a vibration free environment will ensure prolonged service life. Mounting in vibration prone
areasmayintroduceoperationalissues.Pleaseconsidermountinglocationwithcare.
The actual measurement is NOT affected by vibration, it is an instrument life-time reliability
consideration.
Figure8-MountingDimensions
22”
22”
22”