Owner's manual

<3 GENERL SPECIFICATIONS> 3-2
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
Performance specification
Precision* 0.01% 0
2
Linearity* Typically R2 > 0.999
Response time 5 or 10 seconds plus transport time to
analyzer
Drift* Span drift (6-12 month calibration) <+/-
0.1% 0
2
Zero drift (6-12 month Calibration) <+/-
0.05% 0
2
Analog I/O (Optional) Outputs: Concentration/Transmission (3@
4-20mA isolated) Sub 4mA for warnings/
faults
Inputs: Pressure/Temperature Feed for
Compensation (2@ 4-20mA isolated,
powered or loop power)
Digital I/O Outputs: Warning/Fault/Concentration Limit
Relays (3 Form C Relay SPDT rated 1A@
24VDC)
• Valve Control (3@ 24VDC, Max 10W per
valve), zero/span
• Inputs: Remote Validation (3 voltage free
floating contacts) for zero/span
Communications Ethernet, IEEE 802.3, 10/100 Mbps, RJ45
Automatic USB data transfer (upload/
download settings and data)
Calibration Recommended Calibration Check Interval
3-6 Months
Gas Sampling The extracted sample should be typically
Conditions filtered, clean and dry (non-condensing)
Cell volume = 260 cc
Flow rate of 1~20L/min, typically 6 L/min
Pressure of -3 psig to 100 psig
Temperature of -20°C(4°F) to 50°C (122°F
un-heated or 120˚C (248˚F) heated
Gas Measured* O
2
: 0.01% detection limit, Min Range
0-1%, Max range 25%
Performance Specifications are application dependant.
*Consult Yokogawa for ranges; All
performance specifications are for 25ºC at
1 bar.
Installation Specifications
By Design: The analyzer is designed for operation
in General Purpose area. The addition
of a Purge System enables operation in
Hazardous Area for gaseous releases.
Class 1 Division 2 Group B, C and D
(ATEX/CE Pending) (Optional) CSA
Special Acceptance certification.
Flow Cell Wetted Parts
Standard:316L,Sapphirewindows,
TeflonencapsulatedVitonO-rings,and
protected gold mirror
Optional:MonelAlloy400,Kalrez4079
O-rings
Mains voltage: CE/230Vmodel-210to240VAC,50-60
Hz,singlephase110Vmodel-100to
130VAC,50-60Hz,singlephase
Surge and temporary overvoltage
immunity in compliance with
EN61326-1.
Integration
Configuration
Sample is fully extracted from process (and should be
conditioned before measurement)
Process pressure and temperature can be controlled or
the analyzer may require pressure and temperature inputs
(application dependant)
• Length of flow cell is fixed
Purge Gas & Validation
Available systems (standard or custom) for:
• Manual or Automatic Validation
• Manual or Automatic Calibration
• Manual or Automatic Stream Switch
• Analyzer purge gas control
• Other options Available
Display and Software Functions:
TruePeak software has multiple levels, the default (or start page) is
the Main Menu:
Main Menu Displays:
Gas Concentration
Transmission %
Status (warm-up, OK, Warning, Fault, etc.)
Temperature (Fixed, Active Ambient or Active)
Pressure (Fixed or Active)
Main Menu:
Basic Menu
Configure, 3 functions
View Spectra, 2 functions
Data, 3 sub-menus
Trends
Advanced Menu:
Configure, 9 sub-menus (User Password)
Calibrate & Validate, 3 sub-menus
Data, 4 sub-menus
Trends
Active Alarms:
List of active alarms
Shut Down Analyzer:
Instructions to close TruePeak local or VAC
U