Owner manual

IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00
2-1 2-2<2 QUICK START>
6.0
Checking
Ifthereisaninstalledoptional6.5”DisplayandKeypad–ObservetheMainMenumessagesandstatus
information.
6.1 IfthereisnoinstalledUserInterface,thenconnectalaptopPCviaEthernettotheSBCmountedonthe
backplane. Initiate the supplied UltraVNC software from the laptop to initiate a VNC session with the
‘blind’ analyzer and observe the analyzer Main Menu via the laptop.
6.2
AT this time there may be one or more alarm message due to low transmission, out of range parameters or other –
nal system conguration is still required!
6.3 IftheanalyzerdisplaysaWarning“ValidationRequired”thenthisindicatesthatthereisnotargetgas
absorption peak found at start-up. Either, shut down the analyzer, introduce some measured gas into the
owcellandre-startorperformavalidation.Thiswillensurethattheanalyzeriscorrectlytunedtothe
measurement gas absorption peak. If this Warning cannot be cleared by either method, please contact
YokogawaLaserAnalysisDivisionoryourlocalagentforfurtherassistance.
7.0
Congure
BASIC
By way of the appropriate user interface, the correct process parameters and other parameters can now
be entered.
7.1 EntertheBasicMenuandgotoCongure.
7.2
Gas Pressure
Enterinthecorrectprocessgaspressure(ifActive,seeAdvancedCongure).
7.3
Gas Temperature
Enterinthecorrectprocessgastemperature(ifActive,seeAdvancedCongure).
7.4 Ifanyotherparametersarerequiredtobeset(suchasanalogI/Oranges,alarmslevels,AutoValidation
sequences) then the Advanced Menu needs to be accessed.
Advanced Menu access is Password protected (default 1234, can be changed by user if necessary) and should only be
used by skilled and trained persons - Contact Yokogawa Laser Analysis Division or Local Agent if any doubts!
GototheDatasectionunderBasicandconguretheappropriate‘RecordResultData’settings.Thiswill
ensuretheanalyzerstoresimportantinformationduringoperationthatmaybeusedtoverifyoperation/
status/diagnosticsandothertroubleshooting.
8.0
Congure
ADVANCED
Usingthecorrectpassword(Default1234),enterintotheAdvancedMenu,thentheCongure.
8.1 Select the desired measurement units for path length, pressure and temperature.
8.2 Select Fixed or Active.
If Fixed, enter in the correct process gas pressure.
If Active, enter in the 4-20mA input signal range proportional to the pressure transmitter output range.
8.3 Select Fixed or Active.
If Fixed, enter in the correct process gas temperature.
If Active, enter in the 4-20mA input signal range proportional to the temperature range.
8.4 CongurethesystemI/ObyenteringintotheSystemI/OsubmenuinCongure.
8.5 IftheoptionalAnalogI/Oboardisinstalled,thenselectAnalogOutputandsettheappropriate4mAand
20mA values for Ch 1 Concentration and Ch 2 Transmission.
8.6 Selectwhatmode(Block,TrackorHold)the4-20mAoutputsaretobewhentheanalyzerisinWarning,
Fault, Export Data and Calibration Modes.
8.7 CongureDigitalI/O–WarningsandFaults.Manyofthesewillbefactorypre-setsoifunsureaboutany
settings then leave as Factory Default.
SelectandsetlevelforAlarmLimittoeithertheMeasuredGasorTransmission.
8.8 Go to the Data screen and set the appropriate parameters for ‘Record Result Data’ and ‘Spectrum
Capture’. These will ensure the analyzer stores important information during operation that may be used
toverifyoperation/status/diagnosticsandothertroubleshooting.