Owner manual

IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00
3-4 3-5<3 INSTALLATION AND WIRING>
3.4 Purge Gas Requirements and Hazardous Area Systems
TheTDLS220Analyzerrequiresacontinuousnitrogen(optionallyinstrumentair)gaspurgetopreventambient
oxygeningresstotheopticalpath,whenoxygenisthemeasuredgas.Theowratecanbeminimizedaslong
asitpreventsanyambientoxygeningresstothemeasurementopticalpath.Otherpurgegasesmaybeusedas
long as they do not contain any of the measured gas and a clean, dry, etc.
For hazardous area operation, the same nitrogen purge gas is used to purge the entire analyzer (including non-
optical path sections such as the electronics). Refer to purge diagrams below.
Insomeapplications,InstrumentAir(I/A)maybeusedasthepurgegas.Specialsoftwarecongurationsmust
beset-upunderthe“Congure”sectioncalled“Non-ProcessParameters”–refertosoftwaresectionofthis
manual for further details.
Theowratecanbeminimizedaslongasitpreventsanyambientoxygeningresstothemeasurementoptical
path.Otherpurgegasesmaybeusedaslongastheydonotcontainanyof
ELECTRONICS
FLOW CELL
POWER
HEAT
TRACE
LASER &
DETECT
N2 Purge Inlet
N2 Purge Vent
Figure3–SafeArea/GeneralPurposeN2AnalyzerPurgeSchematic