Owner manual

IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00
<2 QUICK START> 2-3
8.9 GototheTrendsscreenandreview/plotseveralofthelistedparameterstocheckanalyzerperformance
over a period of time.
8.10
Non-Process
Parameters
Iftheapplicationistousingpurgegascontainingthetargetgas(e.g.Oxygenmeasurementwith
InstrumentAirPurge)thentheNon-Processparametersshouldbeconguredasdetailedlaterinthis
manual under the Software Section.
9.0
Normal Operation
Whenthesite/eldcongurationiscompleteandtheanalyzerhasoperatedforatleasttwohours
without any functional alarms, then perform an export data routine.
9.1 To Export Data, simply insert an empty USB memory stick in to a USB port on the launch unit back
plane. The data transfer may take several minutes.
DO NOT REMOVE THE MEMORY STICK DURING THIS TIME!
9.2 CloseouttheVNCsoftwareanddisconnecttheservicePC–ifconnected.
9.3 Ensurethedoors/lidsareclosedandtightlysealed.
9.4 The system is now in normal operation mode.
9.5
We RECOMMEND sending all the Exported Data files to Yokogawa Laser Analysis Division along
with any notes and comments. We will then be able to store these files on a master record for
future reference.
Please carefully read the appropriate Sections of this Instruction Manual. The TDLS220 Tunable Diode
Laser (TDL) Analyzer is a technologically advanced instrument that requires the appropriate care when
handling, installing and operating.
Failure to do so may result in damage and can void any warranty!
If there is any doubt about any aspect of the Instrument installation or use, please contact Yokogawa
Laser Analysis Division and/or your authorized Representative/Distributor.