SC42-EP04 cc = 10 cm
-1
2-electrode sensor
This sensor is in the program for historical reasons. It makes little sense to select this sensor for
new applications.
It was used mostly for process water and light process solutions. The operating range can be 1to
100 mS/cmwith higher readings having a very high risk of polarization.The SC42-EP18 or the
SC42-EP08(4-elsensors)coverstheserangeswithmuchgreatercertaintyofavoidingpolariza-
tion.
SC42-EP08 cc = 10 cm
-1
4-electrode sensor
Process applications are where this sensor is used, including light fouling. The best range of ap-
plicationis1mS/cmto500mS/cm.Thefour-electrodesystemcopeswithsomecontaminationfrom
the process, and eliminates the tendency for polarization to occur at the higher conductivity. The
upper limit is determined more often by chemical compatibility. The epoxy material is attacked
by many of the chemicals that are present in high conductivity solutions. It is better suited to flow
throughinstallationsinFS40andFF40adaptersandfittings,butitcanalsobeusedindiptubes-
FD40.
SC42-TP08 (FP08) cc = 10 cm
-1
4-electrode sensor
Thesesensorsareforextremeapplications.ThePTFEorthePVDFusedforthebodyofthesen-
sor gives good chemical resistance for the strongly corrosive applications. These sensors are
intended for usein ranges 10mSto 1000mS/cm. Carefulselection of theholder is needed.The
PVDFsub-assemblyisanobviouschoicefortheSC42-FP08,andthePPflowfittingoftenhassuf-
ficient corrosion resistance for the application. In any case selection must be done with due regard
to the process conditions.
The 2-electrodeversions (SC42-T(F)P04 do exist, but like the epoxy version (SC42-EP04) they
should be discounted for new applications.