User guide

52
IM 12D08N05-01E
Ifitismoreconvenient,theusermaymakesolutionsfromSodiumChloride(NaClorcommon
table salt) with the help of the following relationship table. This table is derived from the IEC norm
60746-3.
Table 11-3. NaCl values at 25 °C
Weight % mg/kg Conductivity
0.001 10 21.4 µS/cm
0.003 30 64.0 µS/cm
0.005 50 106 µS/cm
0.01 100 210 µS/cm
0.03 300 617 µS/cm
0.05 500 1.03 mS/cm
0.1 1000 1.99 mS/cm
0.3 3000 5.69 mS/cm
0.5 5000 9.48 mS/cm
1 10000 17.6 mS/cm
3 30000 48.6 mS/cm
5 50000 81.0 mS/cm
10 100000 140 mS/cm
Note!
Forresistivitymeasurementthestandardresistivityunitsofthecalibrationsolutioncanbecalcu-
lated as follows:
R=1000/G(kΩ•cmifG=µS/cm)
Example:0.001%weight R=1000/21.4=46.7kΩ•cm