User guide
27
IM 12D08N05-01E
5
MENU STRUCTURECOMMISSIONING
Menu Parameter Default Range
values min. max.
PID-controlS1 Setpoint 250.0µS/cm -inf +inf
PID-controlS1 Range 50.0µS/cm -inf +inf
PID-controlS1 ManualReset 0% 0% 100%
PID-controlS1 I-time 3600s 1s 3600s
PID-controlS1 D-time 0s 0s 60s
Duty cycle DC period time 10 s 1 s 1800 s
Pulse freq. Max. pulse freq. 70 p/min 1 p/m 120 p/m
mA1 (simulate) Expire time 0.0 s 0 s 1800 s
AlarmS1 Setpoint 450.0µS/cm(high) -inf +inf
AlarmS2 Setpoint 50.0µS/cm(low) -inf +inf
AlarmS1 Hysteresis 4.500µS/cm 0mS/cm +inf
AlarmS1 DelayTime 0.2s 0s +inf
AlarmS1 ExpireTime 0.0.s 0s 1800s
Hold FixedvaluemA1 12mA 3.6mA 21mA
Hold FixedvaluemA2 12mA 3.6mA 21mA
S2, S3, S4 Similar structure to S1
Enter
PI-control S1
Range
(proportional)
Direction
Setpoint
I-time
250.0μS/cm
50.00μS/cm
Reverse
3600 s
Alarm
Control
Hold
USP
Fail
Off
Simulate