MENU STRUCTURECOMMISSIONING
S2, S3, S4 Similar structure to S1
Menu Parameter Default Range
values min. max.
PID-controlS1 Setpoint 7pH -inf +inf
PID-controlS1 Range 1.00pH 0.001pH +inf
P(ID)-controlS1 ManualReset 0% 0% 100%
PI(D)-controlS1 I-time 3600sec. 1sec. 3600sec.
P(I)D-controlS1 D-time 0sec. 0sec. 60sec.
Duty cycle DC period time 10 sec. 1 sec. 1800 sec.
Pulse freq. Max. pulse freq. 70 p/min. 1 p/min. 120 p/min.
mA1 (simulate) Expire time 0.0 sec. 0 sec. 1800 sec.
AlarmS1 Setpoint 13pH(high) -inf +inf
AlarmS2 Setpoint 1pH(low) -inf +inf
Alarm Hysteresis 0.10 pH 0 pH +inf
Alarm Delay time 0.2 sec. 0 sec. 1800 sec.
Hold FixedvaluemA1 12mA 3.6mA 21mA
Hold FixedvaluemA2 12mA 3.6mA 21mA