Owner manual

27
IM 12B07C05-01E
5
MENU STRUCTURECOMMISSIONING
Menu Parameter Default Range
values min. max.
Limitsandtiming Zerohigh 120mV 0mV 532.4mV
(relative to I.T.P.) 2.03 pH 0 pH 9 pH
Limitsandtiming. Zerolow -120mV -532.4mV 0mV
(relative to I.T.P.) -2.03 pH 9 pH 0 pH
Limitsandtiming Slopehigh 110% 100% 110%
Limitsandtiming Slopelow 70% 70% 100%
Limitsandtiming Stabilizationtime 5sec. 2sec. 30sec.
Limitsandtiming Calib.interval 250days 1day 250days
Buffers Buffertable1,2,3 NIST/DIN19266 Seeappendix1
Zero/Slope/ITP Zero 0mV Zerolow Zerohigh
7 pH Zero low Zero high
Zero/Slope/ITP Slope 100% Slopelow Slopehigh
59.16mV/pH Slopelow Slopehigh
Zero/Slope/ITP ITP 7pH 0pH 14pH
Input1(or2)Impedance Highlimit 200000Ω 1000Ω 1000000Ω
Input1(or2)Impedance Lowlimit 1000Ω 1000Ω 1000000Ω