Owner manual
19
IM 12B07C05-01E
4
OPERATIONOFEXAxtPH450G
4-1. Main display functions
T ag:EXAxt PH45 0
HOLD
W AS H
25.0
236.0
mV
S4 S3 S2 S1
Go to trend screen
Go to zoom screen
Go to status screen
Go to maintenance screen
Figure 4-1. Main Display
A heart “ ” mark is lit on the right-upper corner of the
screenwhenHARTcommunicationisactive.An“Χ”
mark is lit when HART communication is abnormal.
Nothing appears when HART communication is not
used.
Notethatthe“Χ”markmayappearduetooutput
signal noise or the like even when HART communica-
tion is not used. There is not problem when HART
communication is not used. Continue operation while
ignoring the mark.
4-2. Trending graphics
Pressing the button changes the display
into a graphical mode in which the average
measured value is shown on a time scale. The
“Live”valueisalsodigitallydisplayedinatext
box. The time scale ( X-axis) and the primary
valuescale(Y-axis)aresetinthe“DISPLAY
SETUP”menu.Thefullscreendisplaysatrend
of 51 points that represent the average of the
selected time interval. The converter samples
the measurement every second. The trending
graphic also shows the maximum and minimum
measured value in that interval.
Forexampleifthetimescaleissetto4hours,
then the trend is shown for 4 hours prior to the
actual measurement. Each point on the trend
linerepresentstheaverageover4*60*60/51=
282 measurements (seconds).
Maximum
Minimum
Minimum
Maximum
Average
Live reading
9.10 pH
4
Figure 4-2. Trend screen
4-3. Zoom in on details
This button gives access to the diagnostic
information of the converter. The following
messages will appear under normal (default)
conditions:
- Home key back to
mainscreen.
-Onelevelup.
-Scrollchoices
(grey
means deactivated).
- Enter selected data
or choice.
Firstzoomscreen
gives you inside
into the parameters
involving current
measurement. All
following zoom
screens give
additional information
about the device and
lead to logbook data.
Next
Next
Next
Next
Figure 4-3. Detail screen
4-3-1. Actual mA1
=thecurrentoutputinmAofthefirstcurrent
output, which is defined as mA1. The range
and function of this mA output can be set in
Commissioning >> Output setup>> mA1
4-3-2. Actual mA2
=thecurrentoutputinmAofthesecond
current output, which is defined as mA2. The
range and function of this mA output can be
set in
Commissioning >> output setup >> mA2
4-3-3. Zero
=calibratedsensoroffsetinmV.Theoretically
the sensor reads 0 mV in a buffer solution with
the same pH value as the Iso po ten tial pH value
4. OPERATION OF EXAxt PH450G
T
--
>
pH
--
>