Owner manual

10
IM 12B07C05-01E
3-3-5. Grounding the housing
Forthesafetyoftheuserandtoprotectthe
instrument against interference, the housing
must always be connected to ground. This has
to be done by a large area conductor. This
cable can be fixed to the rear of the housing or
by using the internal ground connections using
abraidedwirecable.Seefigure3-8.
The minimum cross sectional area of the pro-
tective grounding wire should be 0.75 mm
2
.
3-3-6. Switching on the instrument
After all connections are made and checked,
the power can be switched on from the power
supply.MakesuretheLCDdisplaycomeson.
After a brief interval, the display will change to
the measured value. If errors are displayed or
a valid measured value is not shown, consult
the troubleshooting section (Chapter 8) before
calling Yokogawa.
DANGER
Figure 3-7. Input and output connections
2 1
- +
DC
POWER
12-24 V /10 W
FUSE: 2A/250 VAC/T
2 1
N L
AC
POWER
100-240 VAC/15 VA/ 50/60Hz
FUSE: 500 mA/250 VAC/T
Fusereplacementshouldbeperformedonlyby
a qualified service personnel.
SeeSec.7.MAINTENANCE,Fuse.
Fuse ratings:
Powersupply Fusetype
12-24VDC,10Wmax 2A/250V,Slow
100-240VAC,15VAmax0.5A/250V,Slow
3-3-2. Access to terminal and cable entry
Terminals 1 and 2 are used for the power sup-
ply. Guide the power cables through the gland
closed to the power supply terminals. The ter-
minals will accept wires of 2.5 mm
2
(14 AWG).
Always use cable finishings if possible.
WARNING
3-3-3. AC power
ConnectterminalLtothephaselineoftheAC
powerandterminalNtothezeroline.Seefig-
ure 3-8 for the power ground. This is separated
from input ground by a galvanic isolation.
3-3-4. DC power
Connect terminal 1 to the positive outlet and
terminal 2 to the negative outlet.
This is separated from input ground by a
galvanic isolation. The size of conductors
should be at least 1.25 mm
2
. The overall cable
diametershouldbebetween6&12mm.