3 INSTALLATIONANDWIRING
3-1. Installation and dimensions
3-1-1. Installation site
The EXAxt 450 converter is weatherproof and
can be installed inside or outside. It should,
however, be installed as close as possible to
the sensor to avoid long cable runs between
sensor and converter. In any case, the cable
length should not exceed 50 metres (162 feet).
Selectaninstallationsitewhere:
• Mechanicalvibrationsandshocksare
negligible
• Norelay/powerswitchesareinthedirect
environment
• Accessispossibletothecableglands
(see figure 3-1)
• Theconverterisnotmountedindirect
sunlight or severe weather conditions
• Maintenanceproceduresarepossible
(avoiding corrosive environments)
The ambient temperature and humidity of
the installation environment must be within
the limits of the instrument specifications.
(Seechapter2).
3-1-2. Mounting methods
Refer to figures 3-2 and 3-3. Note that the
EXAxt converter has universal mounting capa-
bilities:
• Panelmountingusingoptionalbrackets
• Surfacemountingonaplate(usingbolts
from the back)
• Wallmountingonabracket(forexample,
on a solid wall)
• Pipemountingusingabracketona
horizontal or vertical pipe
Sizenominal50A
3. INSTALLATION AND WIRING
Figure 3-1. Housing dimensions and layout of glands
220(8.66")184(7.2")
72(2.8")
A
D E F
B C
A : For output signal
B : For contact input
C : For sensor cable
D : For contact output (S1 and S2)
E : For contact output (S3 and S4)
F : For power supply
Adapter
49
(1.93")
G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)
M20 screw (/AM20)
Approx.
55(2.2")
144(5.67")
144(5.67")
20
(0.79")
20
(0.79")
121.5(4.78")
24.5(1")
Adapter for Coduit Work (optional)
(option code : / AFTG, / ANSI, / AM20)
M20 cable gland
(When shipped, not installed)
Grounding terminal (M4 screw)
Hood (Option code: /H5, some cutout on the left side cover)
Unit: mm (inch)