Owner manual
3
IM 12B07C05-01E
2
GENERALSPECIFICATIONS
F) Contact input : Remote wash cycle start.
G) Temperature compensation
Function :Automaticormanual
Compensation to Nernst equation
Process compensation by configurable temperature coefficient,
NEN6411 for water or strong acids/bases or programmable matrix
H) Calibration :Semi-automatic1or2pointcalibrationusingpre-configuredNIST,
US,DINbuffertables4,7&9,orwithuserdefinedbuffertables,with
automatic stability check
Manual adjustment to grab sample
I) Logbook :Softwarerecordofimportanteventsanddiagnosticdatareadily
available in the display or through HART®.
J) Display :GraphicalQuarterVGA(320x240pixels)LCDwithLEDbacklightand
touchscreen.PlainlanguagemessagesinEnglish,German,French,
Spanish,Italian,Swedish,Portuguese and Japanese.
K) Shipping details
Packagesize :290x300x290mm(LxWxD)(11.5x11.8x11.5inch)
Package weight : app. 2.5 kg (5.5 lbs)
Converter weight : app. 1.5 kg
L) Housing : Cast aluminum housing with chemically resistant coating;
Polycarbonate cover with Polycarbonate flexible window
: ProtectionIP66/NEMA4X/CSAType3S
Colour : Silvergrey
PH450G-A(D)-A : IP66 cable glands are supplied with the unit
PH450G-A(D)-U :NEMA4Xcloseupplugsaremountedintheunusedcableentryholes
and can be replaced by conduit fittings as required
Pipe, Panel or Wall mounting using optional hardware
Optionalconduitadapter
G1/2, 1/2NPT or M20 female
M) Power supply
PH450G-A : Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC
Ratings;50/60Hz Acceptablerange;50Hz±5%,60Hz±5%
Power Consumption; 15 VA
PH450G-D : Ratings; 12-24 V DC Acceptable range; 10.8 to 26.4 V DC
Power Consumption; 10 W
N) Safety and EMC conforming standards
,
Safety : EN61010-1
CSAC22.2No.61010-1
UL61010-1
FM3611ClassI,Div.2,GroupABCD,T6forTa-20to55ºC
EMC : EN61326-1ClassA,Table2(Foruseinindustriallocations)
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
KoreaElectromagneticConformityStandard
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.