Owner manual

3
IM 12B07C05-01E
2
GENERALSPECIFICATIONS
F) Contact input : Remote wash cycle start.
G) Temperature compensation
Function :Automaticormanual
Compensation to Nernst equation
Process compensation by configurable temperature coefficient,
NEN6411 for water or strong acids/bases or programmable matrix
H) Calibration :Semi-automatic1or2pointcalibrationusingpre-configuredNIST,
US,DINbuffertables4,7&9,orwithuserdefinedbuffertables,with
automatic stability check
Manual adjustment to grab sample
I) Logbook :Softwarerecordofimportanteventsanddiagnosticdatareadily
available in the display or through HART®.
J) Display :GraphicalQuarterVGA(320x240pixels)LCDwithLEDbacklightand
touchscreen.PlainlanguagemessagesinEnglish,German,French,
Spanish,Italian,Swedish,Portuguese and Japanese.
K) Shipping details
Packagesize :290x300x290mm(LxWxD)(11.5x11.8x11.5inch)
Package weight : app. 2.5 kg (5.5 lbs)
Converter weight : app. 1.5 kg
L) Housing : Cast aluminum housing with chemically resistant coating;
Polycarbonate cover with Polycarbonate flexible window
: ProtectionIP66/NEMA4X/CSAType3S
Colour : Silvergrey
PH450G-A(D)-A : IP66 cable glands are supplied with the unit
PH450G-A(D)-U :NEMA4Xcloseupplugsaremountedintheunusedcableentryholes
and can be replaced by conduit fittings as required
Pipe, Panel or Wall mounting using optional hardware
Optionalconduitadapter
G1/2, 1/2NPT or M20 female
M) Power supply
PH450G-A : Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC
Ratings;50/60Hz Acceptablerange;50Hz±5%,60Hz±5%
Power Consumption; 15 VA
PH450G-D : Ratings; 12-24 V DC Acceptable range; 10.8 to 26.4 V DC
Power Consumption; 10 W
N) Safety and EMC conforming standards
,
Safety : EN61010-1
CSAC22.2No.61010-1
UL61010-1
FM3611ClassI,Div.2,GroupABCD,T6forTa-20to55ºC
EMC : EN61326-1ClassA,Table2(Foruseinindustriallocations)
EN 61326-2-3
EN 61000-3-2 Class A
EN 61000-3-3
KoreaElectromagneticConformityStandard
Installation altitude: 2000 m or less
Category based on IEC 61010: II (Note)
Pollution degree based on IEC 61010: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other
inclusions which may reduce dielectric strength. Degree 2 is the normal indoor
environment.