CONFIGURATION CHECKLIST FOR SC202
Primary choices default alternatives reference on page menu
Measurement Conductivity Resistivity 5.8- 5.9 SC 01
Range 0-1000 μS/cm max. 1999 mS/cm 5.3 “range”
Temperature unit Celsius Fahrenheit 5.10- 5.11 SC 11
Sensor
Cell constant 0.1 /cm any value between 0.08
and 50
5.8-5.9, 6.1- 6.3 SC 03
Sensor type 2-electrode 4- electrode 5.8- 5.9 SC 02
Temperature compensator Pt1000 Ni100, Pt100, 8k55, Pb36 5.10-5.11 SC 10
Choices
Communication enabled disable HART
(R)
, PH201*B 5.19 SC 60- 62
Burn out inactive HI or LO output on fail 5.14- 5.15 SC 32
Temperature compensation NaCl in water fixed T.C., matrix 5.12, 5.13, 5.5 SC 20- 28; “temp”
USP functionality inactive Fail if USP limits are 9.1, 9.2, 5.17 SC 57
exceeded
HOLD during maintenance inactive HOLD last value or fixed
value
5.17, 5.3- 5.4 “hold”, SC 50
Calibration temperature inactive adjustment +/- 15 °C 5.11 SC 12
ZERO calibration inactive adjustment +/-1 S/cm 5.9 SC 04
Diagnostics hard alarm on hard or soft choices 5.17 SC 53
all errors
Cell fouling alarm active except E13 inactive 5.9 SC 05
Password protection inactive password for different levels
5.17 SC 52
Output in Concentration units inactive linearization of output, w% 5.14 - 5.17 SC 31/35/55
on LCD