Owner's manual
IM12B07D02-01E
Parameter setting 5-19
5-3-5. User interface
Code 50  *RET.   When Auto return is enabled, the transmitter reverts to the measuring mode 
from anywhere in the configuration menus, when no button is pressed during 
the set time interval of 10 minutes.
Code 51  *MODE   The manual impedance check (on demand) can be setup for operation in the 
maintenance mode. (Through the closed front cover).
Code 52  *PASS   Passcodes can be set on any or all of the access levels, to restrict access to 
the instrument configuration.
Code 53  *Err.4.1  Error message configuration. Two different types of failure mode can be set.
 Hard fail gives a steady FAIL flag in the display, A Fail signal is transmitted on 
the outputs when enabled in code 32.
 Soft fail gives a flashing FAIL flag in the display. The call for maintenance is 
a good example of where a SOFT fail is useful. A warning that the regular 
maintenance is due, should not be used to shut down the whole measurement.
Code 54    Not used
Code 55  *CALL.M   Call for maintenance is a trigger to signal that the system has been in service 
for longer than the set time without calibration. The user can set up to 250 days 
as a routine service interval.
Code 56  *DISP   The display resolution can be set to either 0.01pH or 0.1pH. Not applicable to 
the ORP (mV) display.