11-4-4.  Setting of errors (FAIL contact)
Code 53 *Err.4.1-*Err .16 (9 items)
     These set how errors are notified to the user. Either of two modes can be set. "Hard 
Fail" mode lights the "FAIL" indicator on the LCD panel. When "enable status output 
contact" is set for the PH201G distributor (Style B), the FAIL contact remains closed and 
the "wash" contact stops operating. If Burnout functions are enabled by Code 32, then 
the output becomes Burnout Upscale or Burnout Downscale.       
"Soft Fail" mode blinks the "FAIL" indicator on the LCD panel. When "enable status 
output contact" is set for the PH201G distributor (Style B), the FAIL contact pulses and 
the "wash" contact and maintenance contact continue to operate normally. In "Soft Fail" 
mode, notification that "periodic maintenance is due" still functions (service code 55) so 
– for this reason – it is better to set "Soft Fail" mode than "Hard Fail" mode.
T5.3.5E.eps
Code Display  Function  Function detail X Y   Z Ddefault values
User interface
53  *Err.4.1  Error setting  Impedance low (input 1)  Soft fail  0
      Impedance low (input 1)  Hard fail  1      1  Hard
  *Err.5.1    Impedance higt (input 1)  Soft fail  0
      Impedance high (input 1)  Hard fail  1      1  Hard
  *Err.4.2    Impedance low (input 2)  Soft fail  0
      Impedance low (input 2)  Hard fail  1      1  Hard
  *Err.5.2    Impedance higt (input 2)  Soft fail  0
      Impedance high (input 2)  Hard fail  1      1  Hard
  *Err.07    Temperrature too high  Soft fail  0
      Temperrature too high  Hard fail  1      1  Hard
  *Err.08    Temperrature too low  Soft fail  0
      Temperrature too low  Hard fail  1      1  Hard
  *Err.09    pH out of range  Soft fail  0
      pH out of range  Hard fail  1      1  Hard
  *Err.11    Wash recovery check  Soft fail  0      0  Soft
      Wash recovery check  Hard fail  1
  *Err.16    Call for maintenance  Soft fail  0      0  Soft
      Call for maintenance  Hard fail  1