12-4. Quick reference for parameter setting
Preparation for operation entails deciding maintenance-mode settings and measurement range set-
ting-related issues. Service level (mode) settings are described in Sec. 5-3. Shipping-time measure-
ment parameter setting defaults are shown in Table 12-1. In general there is no need to change these
defaults.
Table 12-1 Default values of parameters.
T6.1E.eps
25°C
Linear
0 to 1000mS
0 to 100%
20
31
35
1.88
ON(1)
-
OFF(0)
OFF
NaCl
0
-
0.0.0
3
50
12
32
55
21
22
23-28
52
70
Code
Cell constant
Auto return functions
Calibrate temperature
"Burnout" on fail
Display in Weight%
Temp. coefficient
Matrix temp. compensation
Compensation matrix
Passwords
Revert to factory defaults
Corresponding Parameter Default When you should change Refer to
Reference temperature
Output characteristic
Output table
To use different reference temp. Sec. 5-3-3
For non-linear characteristic Sec. 5-3-4
For non-linear characteristic Sec. 5-3-4
When detector at least 30mm from metal pipe walls Sec. 5-3-1
Usually corrected by calibration. Seldom
nevessary to enter value by service code 37
To lock HOLD Sec. 5-3-5
To calibrate against more accurate thermometer Sec. 5-3-2
(esp. when using one-point calibration)
Can set output "burnout" upscale/downscale Sec. 5-3-4
if abnormal
To display concentration units Sec. 5-3-5
Temp. coefficient other than NaCl
Complex compensation Sec. 5.3.3
Can be user defined
To protect settings Sec. 5-3-5
To revert to factory defaults Sec. 5-3-7