Owner's manual

IM 12D06A03-01E
Appendix 1-3
Primary choices Defaults Alternatives Reference Menu or
on page Service code (SC)
Measurement Conductivity
1st Temp. compensation NaCl in water Fixed T.C., Matrix 5-2-4 ‘Commissioning’
2nd Temp. compensation NaCl in water Fixed T.C., Matrix (inactive) 5-2-5 ‘Commissioning’
2nd Line display Process temp. 1st compensation method, Weight%, 5-1-3/4 ‘Maintenance’
Output (mA), C.C., Ref. Temp., Software
release, 2nd compensated Conductivity,
2nd compensation method
Range (Linear) 0-1000 mS/cm 0-1999 mS/cm 5-2-1 ‘Commissioning’
Temperature compensator 30k NTC Pt 1000 5-3-2 SC 10
Temperature unit Sensor Celsius (ºC) Fahrenheit ( ºF ) 5-3-2 SC 11
Cell constant 1.88 /cm Any value between 0.2 and 19.99 /cm 5-3-1 SC 03
Communication HART® enabled disable HART®, PH201*B 5-7 SC 60
Burn out inactive HI or LO, Pulse burnout 5-5 SC 32
HOLD during maintenance inactive Hold last value or fixed value 5-2-2 ‘Commissioning’
Calibration temperature inactive Adjustment +/- 15ºC 5-3-2 SC 12
AIR (zero) calibration inactive Adjustment +/- 15 S/cm 5-3-1 SC 04
C.C. Calibration inactive 0.2 and 19.99 /cm 5-1-2 ‘Maintenance’
Diagnostics Hard alarm (all errors) Hard or soft choices 5-6 SC 53
Instrument Limit: E5 (high) 3S E5 (High) 0.10 - 9.99 S 5-6 SC 54
Instrument Limit: E6 (low) 0.5 S E6 (Low) 0.00 - 99.9 S 5.6 SC 54
Password protection inactive (In)active password for different levels 5-6 SC 52
Output in Concentration units inactive Linearization of output, Weight% on LCD 5-6 SC 55
Code Display Default Values Setting #1 Setting #2 Setting #3
03 *C.C. 1.88 /cm
10 *T.SENS 0 : 30 kNTC
11 *T.UNIT 0 : ºC
12 *T.ADJ 0 ºC
20 *T.R. ºC 25ºC
21 *T.C.1 2.1% per ºC
*T.C.2 2.1% per ºC
22 *MATRX 1 : H2SO4
31 *OUTP.F 0: Linear
32 *BURN 0: No Burnout
50 *RET 1: On
52 *PASS 0.0.0.: off.off.off
53 *Err. 05 1 : HARD
*Err. 06 1 : HARD
*Err. 07 1 : HARD
*Err. 08 1 : HARD
54 *E5.L 3S
*E6.LIM 5S
*OUTP 0% : 0S
100% : 1000mS
*HOLD disabled
*TEMP.1 NaCl
*TEMP.2 NaCl
60 *COMM. 1.0: On, write enable
mAmA
mAmA
mAmA
mAmA
11-5. Coded service settings (default)
11-4. Configuration Checklist For ISC202