Owner's manual

IM 12D06A03-01E
6-2 Calibration
ENT
NOYES
MODE
MEASURE
CAL
DISPLAY
HOLD
MODE
6-2. Calibration procedure
Press the MODE key
The legend CALIB appears,
and the YES/NO key prompt flags flash
Put the sensor in standard solution
Press YES
Set the value using the key
Select the flashing digit with the key
Increase it’s value by pressing the key
When the correct value is displayed,
press ENT to enter the change
After briefly displaying WAIT,
the CAL.END message appears
The calibration is now complete
Put the sensor back in the process and press YES
The cell constant is automatically updated after the calibration and the new value can be read on the dis-
lay as described in section 5-1-2. and 5-1-3.
If the calibrated cell constant is not within the range of 0.2 - 19.99 cm
-1
, error E3 is displayed.