Parameter setting 5-15
Code Display Function Function detail X Y Z Default values
Temperature compensation functions
20 *T.R.°C Set reference temp. Use
keys to set value 25 °C
21 *T.C.1 Set temp. coef. 1 Adjust compensation factor 2.10 %
if set to *T.C. in section 5-2-5. per °C
Set value with keys
*T.C.2 Set temp. coef. 2 Adjust compensation factor 2.10 %
if set to *T.C. in section 5-2-5. per °C
Set value with keys
22 *MATRx Select matrix Choose matrix if set to *MATRX
in section 5-2-5, using keys
H
2
SO
4
, 0 -100ºC, 0.5 - 5% 1 1 H
2
SO
4
H
2
SO
4
, 0 -100ºC, 2.5 - 25% 2
HCl, 0 - 60ºC, 0.5 - 5% 3
HCl, 0 - 60ºC, 1 - 20% 4
HNO
3
, 0 - 80ºC, 0.5 - 5% 5
HNO
3
, 0 -80ºC, 2.5 - 25% 6
NaOH, 0 -100ºC, 0.5 - 5% 7
NaOH, 0 -100ºC, 0.5 - 15% 8
User programmable matrix 9
23 *T1 °C (°F) Set temp. range Enter 1st (lowest) matrix temp. value
*T2.. Enter 2nd matrix temp. value
*T3.. Enter 3rd matrix temp. value
*T4.. Enter 4th matrix temp. value
*T5.. Enter 5th (highest) matrix temp. value
24 *L1xT1 Enter conductivity Value for T1
*L1xT2 values for lowest Value for T2
.... concentration
*L1xT5 Value for T5
25 *L2xT1 Concentration 2 Similar to code 24
26 *L3xT1 Concentration 3 Similar to code 24
27 *L4xT1 Concentration 4 Similar to code 24
28 *L5xT1 Concentration 5 Similar to code 24
29 Not used