2-8 Specifications
2-4. Control Drawing ISC202S mA HART
®
Specification (ATEX)
ISC202S
(Inductive Conductivity transmitter)
Safe area
Uo = 31.5 Volt DC
Io = 100 mA
Po = 1.2 Watt
EEx ia or ib Certified Repeater
Power Supply
(HART Compatible)
Output
Supply
ISC40S Sensor
terminals 11-17
(KEMA 00ATEX1067 X)
+
_
G
ISC40S Sensor
terminals 11-17
(KEMA 00ATEX1067 X)
Hazardous area Safe area
+
_
G
Functional
earth
Functional
earth
Intrinsically safe design
CENELEC standard EEx ia IIC: T4 for ambient temp. < 55°C
T6 for ambient temp. < 40°C
Certificate nr. KEMA 06ATEX0222 X
ISC202S
+
_
Load
Resistance
EEx ia or ib
Certified safety barrier or power
with Rint=300
:
Io = 100 mA
Uo = 31.5 Volt DC
Hazardous area
Functional
earth
+
_
24 volts DC Nominal
Supply Voltage.
Intrinsically safe design
CENELEC standard EEx ia IIC: T4 for ambient temp. < 55°C
T6 for ambient temp.< 40°C
Certificate nr. KEMA 06ATEX0222 X
(HART compatible)
(Indutive Conductivity transmitter)
Zone 0 or 1
Zone 0 or 1
・ Electrical data of the ISC202S.
- Supply and output circuit (terminals + and -):
Maximum input voltage U
i
= 31.5 V. Maximum input current I
i
= 100 mA.
Maximum input power P
i
= 1.2 W.
Effective internal capacitance C
i
= 22 nF.
Effective internal inductance L
i
= 35 PH.
- Sensor input circuit (terminals 11 through 17):
Maximum output voltage U
o
= 14.4 V. Maximum output current I
o
= 20 mA.
Maximum allowed external capacitance C
o
= 600 nF (for ISC202S-A),
C
o
= 3.5 PF (for ISC202S-N)
Maximum allowed external inductance L
o
= 88 mH (for ISC202S-A),
L
o
= 200mH (for ISC202S-N)
・ Barriers and power supply specification must not exceed the maximum values as
shown in the diagram above. These safety descriptions cover most of the commonly
used industry standard barriers, isolators and power supplies.
・ The Hand Held Communicator must be of a ATEX certified intrinsically safe type in
case it is used on the intrinsically safe circuit in the hazardous area or of a ATEX
certified non-incendive type in case it is used in the non-incendive circuit in the
hazardous area.