Maintenance 7-2
7-3. Fuse Replacement
This section describes how to replace the power supply fuse.
Use only a fuse of the specified current, voltage and type ratings to prevent fire.
Power to the instrument must be turned off at the external breaker before replacing the fuse. After
replacing the fuse, install a high-voltage shield plate before turning the power on.
Do not short circuit the fuse holder.
Specified Ratings of the Power Supply Fuse
The power supply fuse (recommended fuse) used in the DC402G is as follows.
115 V AC version
Maximum rated voltage: 250 V
Maximum rated current: 200 mA
Compliance: UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law
Part number: A1105EF
230 V AC version
Maximum rated voltage: 250 V
Maximum rated current: 100 mA
Compliance: UL, CSA, VDE, Japan’s Electrical Appliance and material Safety Law
Part number: A1103EF
How to replace the fuse
1) Before replacing the fuse, turn off power to the instrument at the external breaker.
2) Remove the instrument cover and the high-voltage shield plate.
3) Remove the cover from the fuse holder by pulling out by hand.
4) Remove the fuse and install a new, recommended or equivalent fuse on the holder.
5) Put the fuse cover back on the holder securely.
6) Install the high-voltage shield plate.
It is recommended that the fuse be replaced every two years even if it has not blown.
Shield Plate
Fuse Cover
Fuse Holder
Figure 7-1. How to replace the fuse
DANGER