User guide

IM 12D08E02-01E
Calibration 6-3
6-3. Calibration with HOLD active
mS/cm
HOLD
mS/cm
ENT
HOLD
mS/cm
ENT
HOLD
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
mS/cm
NOYES
HOLD
YES
mS/cm
NOYES
HOLD
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
mS/cm
NOYES
HOLD
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
mS/cm
NO
YES
HOLD
mS/cm
NOYES
NO
HOLD
NO MODEYES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
mS/cm
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
MEASURE
DISPLAY 1
HOLD
DISPLAY 2
CAL 2
CAL 1
Note: See section 4-5 for full
display access function