6-2 Calibration
6-2. Calibration procedure
ENT
M Ω.cm
ENT
M Ω.cm
>
Select the flashing digit with the > key.
Increase its value by pressing the key
>
Set the value
using the >, , ENT key.
NOYES
M Ω.cm
NOYES
M Ω.cm
Put the sensor in standard
solution. Press YES.
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES to return to the measuring mode.
ENT
M Ω.cm
M Ω.cm
YES
NO
YES
M Ω.cm
NO
NO MODEYES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
M Ω.cm
CAL 2
CAL 1
MEASURE
DISPLAY 1
HOLD
DISPLAY 2
Press the MODE key.
The legend CAL.1/CAL.2
appears, and the YES/NO
key prompt flags flash.
If the SETP legend
appears, press NO first.
MODE
Note: See section 4-5 for full
display access details