User guide
IM 12D08E02-01E
Parameter setting 5-25
Code Display Function Function detail X Y Z Default values
User interface
50 *RET Auto return Auto return to measuring mode Off 0
Auto return to measuring mode On 1 1 On
51 *MODE Mode setup Setpoints in maintenance mode Off 0 0 Off
Setpoints in maintenance mode On 1
52 *PASS Passcode Maintenance passcode Off 0 0.0.0 Off
Note # = 0 - 9, where Maintenance passcode On #
0 = no passcode Commissioning passcode Off 0 Off
1=111, 2=333, 3=777 Commissioning passcode On #
4=888, 5=123, 6=957 Service passcode Off 0 Off
7=331, 8=546, 9=847 Service passcode On #
53 *Err. 1. 1(2) Error setting Polarisation too high Soft fail 0
Polarisation too high Hard fail 1 1 Hard
*Err. 5. 1(2) Shorted measurement Soft fail 0
Shorted measurement Hard fail 1 1 Hard
*Err. 6. 1(2) Open measurement Soft fail 0
Open measurement Hard fail 1 1 Hard
*Err. 7. 1(2) Temperature sensor open Soft fail 0
Temperature sensor open Hard fail 1 1 Hard
*Err. 8. 1(2) Temp. sensor shorted Soft fail 0
Temp. sensor shorted Hard fail 1 1 Hard
*Err. 13. 1(2) USP <645> limit exceeded Soft fail 0 0 Soft
USP <645> limit exceeded Hard fail 1
*Err.22. 1 Control time-out Soft fail 0 0 Soft
Control time-out Hard fail 1
*SOFT SOFT FAIL CONTACT Pulsing contact 0 0 Pulsing
No Pulsing contact 1
54 *E5.LIM 1(2) E5 limit setting Maximum conductivity value 25 mS
(Minimum resistivity value) 0.04 kΩ
*E6.LIM 1(2) E6 limit setting Minimum conductivity value 1 µS
(Maximum resistivity value) 1 MΩ
55 *% Display mA in w/w% mA1-range displayed in w/w% off 0 0.0 Off
mA1-range displayed in w/w% on 1
mA2-range displayed in w/w% off 0 Off
mA2-range displayed in w/w% on 1
*%1 Set w/w% for range 1 Press YES to access 0% value adjustm.
*0% Set 0% output value in w/w%
*100% Set 100% output value in w/w%
*%2 Set w/w% for range 2 Press YES to access 0% value adjustm.
*0% Set 0% output value in w/w%
*100% Set 100% output value in w/w%
56 *DISP Display resolution Auto ranging display 0 0 Auto
Display fixed to X.XXX µS/cm or MΩ•cm 1
Display fixed to XX.XX µS/cm or MΩ•cm 2
Display fixed to XXX.X µS/cm or MΩ•cm 3
Display fixed to X.XXX mS/cm or kΩ•cm 4
Display fixed to XX.XX mS/cm or kΩ•cm 5
Display fixed to XXX.X mS/cm or kΩ•cm 6
Display fixed to XXXX mS/cm or kΩ•cm 7
57 *USP USP setting Disable the E13 (USP limit passed) 0 0 0.0 Off/Off
X=Cell 1 Y= Cell 2 Enable the E13 (USP limit passed) 1 1
58 *CALC Calculation setting Choose the main parameter for display
Ratio (a/b) 0 5 No calc.
Differential (a-b) 1
% passage b/a 2
% rejection (100 (a-b)/a) 3
Deviation (100 X (b-a)/a) 4
No calculation SC1 in display 5
VGB directive 450L, pH calculation 6
When set for resistivity measuring, only
selection 1 and 5 are available