User guide
IM 12D08E02-01E
Parameter setting 5-21
Code Display Function Function detail X Y Z Default values
Contacts
Contin. Calculated value 0
41 Meas cond/res. value cell 1 1 Cell 1
Meas cond/res. value cell 2 2
Meas temp. value cell 1 3
Meas temp. value cell 2 4
Meas temp. T1-T2 5
42 *S3 Relay 3 settings Off 0 0.0.0 Off
Low setpoint 1
High setpoint 2
“HOLD” active 3
Process alarm 0
Proportional duty cycle control 1
Proportional frequency control 2
PI duty cycle ** 3
PI pulse freq.** 4
USP contact 5
Calculated value 0
Meas cond/res. value cell 1 1
Meas cond/res. value cell 2 2
Meas temp. value cell 1 3
Meas temp. value cell 2 4
Meas temp. T1-T2 5
43 *S4 Relay 4 settings Off 0 4.0.0
Low setpoint 1
High setpoint 2
“HOLD” active 3
Note: “FAIL” relay Fail alarm 4 FAIL
contact is used to Process alarm 0 Alarm
indicate when the Proportional duty cycle control 1
diagnostics detect a Proportional frequency control 2
problem PI duty cycle ** 3
PI pulse freq.** 4
USP contact 5
Calculated value 0
Meas cond/res. value cell 1 1
Meas cond/res. value cell 2 2
Meas temp. value cell 1 3
Meas temp. value cell 2 4
Meas temp. T1-T2 5