    <9.  Parameter Lists>
9-5
IM 01C25T02-01E
Relative 
Index  
Index Parameter Name Factory Default
Write 
Mode
Explanation
119 1119 DEVICE_
CONDITION_
ACTIVE_4
0x00000000 — FD_EXTENDED_ACTIVE_4 & FD_EXTENDED_MAP_4
120 1120 DEVICE_
CONDITION_
ACTIVE_5
0x00000000 — FD_EXTENDED_ACTIVE_5 & FD_EXTENDED_MAP_5
121 1121 DEVICE_
CONDITION_
ACTIVE_6
0x00000000 — FD_EXTENDED_ACTIVE_6 & FD_EXTENDED_MAP_6
122 1122 DEVICE_
CONDITION_
ACTIVE_7
0x00000000 — FD_EXTENDED_ACTIVE_7 & FD_EXTENDED_MAP_7
123 1123 DEVICE_
CONDITION_
ACTIVE_8
0x00000000 — FD_EXTENDED_ACTIVE_8 & FD_EXTENDED_MAP_8
9.2  SENSOR Transducer Block
Relative 
Index 
Index  Parameter Name  Factory Default 
Write 
Mode 
Explanation 
0 2000 Block Header  TAG: “STB” Block 
Tag = 
O/S 
Information on this block such as Block Tag, DD Revision, 
Execution Time etc. 
1 2001 ST_REV  — — The revision level of the static data associated with the function 
block. The revision value will be incremented each time a static 
parameter value in the block is changed. 
2 2002 TAG_DESC  Null  AUTO  The user description of the intended application of the block 
3 2003 STRATEGY  1 AUTO  The strategy eld can be used to identify grouping of blocks. 
This data is not checked or processed by the block. 
4 2004 ALERT_KEY  1 AUTO  The identication number of the plant unit. This information 
may be used in the host for sorting alarms, etc. 
5 2005 MODE_BLK  AUTO  AUTO  The actual, target, permitted, and normal modes of the block. 
6 2006 BLOCK_ERR  — — This parameter reects the error status associated with 
hardware or software components associated with a block. It is 
a bit string, so that multiple errors may be shown. 
7 2007 UPDATE_EVT  — — This alert is generated by any change to the static data. 
8 2008 BLOCK_ALM  — — The block alarm is used for all conguration, hardware, 
connection failure or system problems in the block. The cause 
of the alert is entered in the subcode eld. The rst alert to 
become active will set the Active status in the Status attribute. 
9 2009 TRANSDUCER_ 
DIRECTORY 
— — A directory that species the number and starting indices of the 
transducers. 
10 2010 TRANSDUCER_ 
TYPE 
100 (Standard 
Pre-ssure with 
Calibration) 
— Identies transducer. 
11 2011 XD_ERROR  — — The error code in transducer. 
0=No failure 
18=Calibration error 
20=Electronics failure 
22=I/O failure 
12 2012 COLLECTION_ 
DIRECTORY 
— — A directory that species the number, starting indices, and DD 
Item Ids of the data collections in each transducer within a 
transducer block. 
13 2013 PRIMARY_ VALUE_
TYPE 
107: differential 
pressure 
108: gauge 
pressure 
109: absolute 
pressure 
O/S  The type of measurement represented by primary value. 
Followings are available for the transmitter: 
107=differential pressure 
108=gauge pressure 
109=absolute pressure 
14 2014 PRIMARY_ VALUE  — — The measured value and status available to the function block. 
15 2015
PRIMARY_ VALUE_
RANGE 
Range of capsule  — The High and Low range limit values, engineering units code 
and the number of digits to the right of the decimal point to be 
used to display the primary value. 
16 2016 CAL_POINT_HI  Max range  O/S  The highest calibrated value. 
17 2017 CAL_POINT_LO  0 O/S  The lowest calibrated value. 