User Manual
    <9.  Parameter Lists>
9-3
IM 01C25T02-01E
Relative 
Index  
Index Parameter Name Factory Default
Write 
Mode
Explanation
49 1049 DEVICE_STATUS_5  reserve — Device status For details, refer to Table 8.5 
50 1050 DEVICE_STATUS_6  reserve — Device status For details, refer to Table 8.6 
51 1051 DEVICE_STATUS_7  reserve — Device status For details, refer to Table 8.7 
52 1052 DEVICE_STATUS_8  reserve — Device status For details, refer to Table 8.8 
53 1053 SOFTDWN_
PROTECT 
0x01  AUTO  Denes whether to accept software downloads. 
0x01: Unprotected 
0x02: Protected 
54 1054 SOFTDWN_FORMAT  0x01  AUTO  Selects the software download method. 
0x01: Standard 
0x02: YOKOGAWA Standard
55 1055 SOFTDWN_COUNT  0 — Indicates the number of times the internal FlashROM was 
erased. 
56 1056 SOFTDWN_ACT_ 
AREA 
0 — Indicates the ROM number of the currently working FlashROM. 
0: FlashROM #0 working 
1: FlashROM #1 working 
57 1057 SOFTDWN_MOD_
REV 
1, 0, 0, 0, 0, 0, 0, 
0, 0 
— Indicates the software module revision. 
58 1058 SOFTDWN_ERROR  0 — Indicates the error during a software download. 
Refer Table A7.4. 
59 1059 SOFTDWN_HISTORY — Not used for the transmitter.
60 1060 SOFTDWN_HIST_
INDEX 
0 AUTO Not used for the transmitter.
61 1061 COMPATIBILITY_REV  1 — Indicates the smallest Rev value compatible with device 
DevRev. 
62 1062 CAPABILITY_LEV  0x00 — Indicates the capability level of instrument interior. 
63 1063 CAPABILITY_CONFIG  0x0000 — A parameter corresponding to AP_CONF or DEV_OPTIONS 
before the addition of parameter CAPABILITY_LEV. 
64 1064 WRITE_LOCK_LEVEL  2 AUTO Species blocks that activates Write Lock. 
65 1065 SI_CONTROL_
CODES 
1 — A parameter for switching to make the instrument compatible 
with SI units. 
66 1066 FD_VER  0 — Indicates value of major version of instrument diagnostics 
specications (FF-912). 
67
1067 FD_FAIL_ACTIVE  0 — A parameter that corresponds to ”Failed” in the NAMUR NE-
107 category.
68 1068 FD_OFFSPEC_
ACTIVE 
0 — A parameter that corresponds to ”Off Specication” in the 
NAMUR NE-107 category.
69 1069 FD_MAINT_ACTIVE  0 — A parameter that corresponds to ”Maintenance” in the NAMUR 
NE-107 category.
70 1070 FD_CHECK_ACTIVE  0 — A parameter that corresponds to ”Check Function” in the 
NAMUR NE-107 category.
71 1071 FD_FAIL_MAP  0xFC000000  AUTO Species the bit assigned to FD_FAIL_ACTIVE, a parameter 
for indicating ”Failed,” a 32-bit alarm listed in FD_SIMULATE.
DiagnosticValue.
72 1072 FD_OFFSPEC_MAP  0x00003800  AUTO Species the bit assigned to FD_OFFSPEC_ACTIVE, a 
parameter for indicating ”Off Specication,” a 32-bit alarm 
listedin FD_SIMULATE. DiagnosticValue.
73 1073 FD_MAINT_MAP  0x000003E0  AUTO Species the bit assigned to FD_MAINT_ACTIVE, a 
parameterfor indicating ”Maintenance,” a 32-bit alarm listed in 
FD_SIMULATE. DiagnosticValue.
74 1074 FD_CHECK_MAP  0x01FF8008  AUTO Species the bit assigned to FD_CHECK_ACTIVE, a 
parameter for indicating ”Check Function,” a 32-bit alarm listed 
in FD_SIMULATE. DiagnosticValue.
75 1075 FD_FAIL_MASK  0xFFFFFFFF  AUTO Species the bit that noties the host of 32-bit ”Failed” alarms 
listed in FD_FAIL_ACTIVE.
76 1076 FD_OFFSPEC_MASK  0xFFFFFFFF  AUTO Species the bit that noties the host of 32-bit ”Off 
Specication” alarms listed in FD_OFFSPEC_ACTIVE.
77 1077 FD_MAINT_MASK  0xFFFFFFFF  AUTO A parameter that species the bit that noties the host of 
32-bit ”Maintenance” alarms listed in FD_MAINT_ACTIVE. A 
parameter set by the user. 
78 1078 FD_CHECK_MASK  0xFFFFFFFF  AUTO Species the bit that noties the host of 32-bit ”Check 
Function”alarms listed in FD_CHECK_ACTIVE.
79 1079 FD_FAIL_ALM  AUTO  Indicates alarm information for alarms categorized under 
”Failed.”
80 1080 FD_OFFSPEC_ALM  AUTO  Indicates alarm information for alarms categorized under ”Off 
Specication.”