User Manual
    <6.  Explanation of Basic Items>
6-4
IM 01C25T02-01E
Determination of the range limit of static 
pressure signal:
When the static pressure signal exceeds the 
maximum working pressure (MWP) of the capsule, 
set the status of PRIMARY_VALUE to Uncertain-
Subnormal. Also, set the status of SECONDARY_
VALUE and TERTIARY_VALUE to Uncertain-
Sensor Conversion not Accurate. The status under 
normal conditions is Good-Non Specic.
Adjustment of static pressure signal:
The zero/span adjustment function can be used 
for static signals, as in the case of pressure 
/differential pressure signals. However, the zero-
point adjustment by the external screw function is 
not supported for static pressure signal.
Automatic adjustment
The value of the static pressure exerted on the 
point where adjustment is to be made is assigned 
to parameters (CAL_SP_POINT_LO, CAL_SP_
POINT_HI). After this assignment is set, the amount 
of adjustment is calculated by the transmitter and 
automatically updated. The procedure for adjusting 
the static pressure signal is the same as that of 
pressure/differential pressure signal.
Manual adjustment
From the exerted pressure and the output of the 
transmitter, calculate the amount of zero/span 
adjustment manually and assign the calculated 
value to parameters (CAL_SP_DEVIATION_LO, 
CAL_SP_DEVIATION_HI).
Selection of static pressure signal type:
Differential pressure transmitters measure the 
absolute static pressure. Based on the assumption 
that atmospheric pressure is constant, these 
transmitters can output signal equivalent to gauge 
pressure calculated by deducting the atmospheric 
pressure (xed value) from the measured absolute 
pressure. The type of static pressure signal output 
to SECONDARY_VALUE and TERTIARY_VALUE 
is selected in SP_VALUE_TYPE. SP_VALUE_
TYPE allows selection of gauge pressure or 
absolute pressure. When selecting gauge pressure, 
set the value of the atmospheric pressure (xed 
value) to ATM_PRESS. The default value of 
ATM_PRESS is equal to the value of standard 
atmospheric pressure (101.325 kPa). 
6.3.5  Functions Relating to Capsule and 
Amplier Temperature
Reference to capsule and amplier temperature 
value:
In CAP_TEMP_VAL and AMP_TEMP_VAL, it 
is possible to refer to the capsule and amplier 
temperature value and status. The update period 
of this value is about 1 second. The temperature 
unit is selected by XD_SCALE.unit of the AI block, 
in which CAP_TEMP_VAL is selected. The status 
is normally Good-Non Specic. However, in the 
case of sensor failure or out of measurement range, 
it turns to Bad or Uncertain. For specics, refer to 
Table 8.9.
Determination of the range limit of capsule 
temperature:
The range of capsule temperature is from –40 to  
120°C. When the measured capsule temperature 
deviates from –50 to 130°C range, set the status of 
CAP_TEMP_VAL to Uncertain-Sensor Conversion 
not Accurate. The status under normal conditions 
is Good-Non Specic. Also, set the status of 
PRIMARY_VALUE, SECONDARY_VALUE and 
TERTIARY_VALUE TO Uncertain-Subnormal.
Determination of the range limit of amplier 
temperature:
The range of amplier temperature is from –40 to 
120°C. When the measured amplier temperature 
deviates from –50 to 95°C range, set the status of 
AMP_TEMP_VAL to Uncertain-Sensor Conversion 
not Accurate. The status under normal conditions is 
Good-Non Specic.










