User guide

<5. Modbus Communication>
5-9
IM 01C25R05-01EN
Address
Label used in
DTM
Address Type Access Type
Char/
Date
Read/
Write
Enum list
Default
Value
*1
Unit
Valid
Range
IM
Chapter
1461 ET LRV Holding
Register
FLOAT - W - -200.0 °C -210 to
860
3.3.3
1463 Pres USL Holding
Register
FLOAT - R - 100.0 kPa -
1465 Pres LSL Holding
Register
FLOAT - R - -100.0 kPa -
1467 Pres Min span Holding
Register
FLOAT - R - 0.5 kPa -
1469 SP USL Holding
Register
FLOAT - R - 25.0 MPa -
1471 SP LSL Holding
Register
FLOAT - R - 0.0 MPa -
1473 SP Min span Holding
Register
FLOAT - R - 1.0 MPa -
1475 ET USL Holding
Register
FLOAT - R - 850.0 °C -
1477 ET LSL Holding
Register
FLOAT - R - -200.0 °C -
1479 ET Min Span Holding
Register
FLOAT - R - 10.0 °C -
1551 Pres Damp
Holding
Register
FLOAT - W - 2.0 s 0 to 100 3.3.5
1553 SP Damp Holding
Register
FLOAT - W - 1.0 s 0 to 100 3.3.5
1555 ET Damp Holding
Register
FLOAT - W - 1.0 s 0 to 100 3.3.5
1601 Pres trim Auto
Upper Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
1603 Pres trim Auto
Lower Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
1605 Pres trim
Manual Upper
Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1607 Pres trim
Manual Lower
Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1609 SP trim Auto
Upper Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1611 SP trim Auto
Lower Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1613 SP trim Manual
Upper Pt
Holding
Register
FLOAT - W - -
- With in
adjustable
range
3.4.4
1615 SP trim Manual
Lower Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1617 ET trim Auto
Upper Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1619 ET trim Auto
Lower Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1621 ET trim Manual
Upper Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1623 ET trim Manual
Lower Pt
Holding
Register
FLOAT - W - - - With in
adjustable
range
3.4.4
1625 Trim Clear Code Holding
Register
USIGN8 - W Table 72 0x0000 - 0 to 5 3.4.4
*1: This value is factory default of M range measurement span without order specication.
*2: Upper case character can be used. Lower-case is translated to upper case.