User guide
    <4.  Diagnostics>
4-5
IM 01C25R05-01EN
Integral 
Indicator
DTM 
display
Status 
group
Cause Countermeasure
Value and Status (Data Quality and Limit Status)
Differential 
Pressure 
(DP)
Static 
Pressure 
(SP)
External  
Temperature 
(ET)
DP % SP % ET %
AL.55 
SP. ADJ
SP SPAN 
trim err
7 Specied value 
is outside of 
setting range.
Adjust settings 
and change 
them as 
needed.
Value:
Measured 
value
Status:
Good 
and Not 
Limited
(0xC0)
Value:
Measured 
value
Status:
Poor 
Accuracy 
and Not 
Limited
(0x40)
Value:
Measured 
value
Status:
Good and 
Not Limited
(0xC0)
Value:
Measured 
value
Status:
Good 
and Not 
Limited
(0xC0)
Value:
Measured 
value
Status:
Poor 
Accuracy 
and Not 
Limited
(0x40)
Value:
Measured 
value
Status:
Good 
and Not 
Limited
(0xC0)
SP ZERO 
trim err
AL.56 
ET. RNG
Illegal ET 
LRV
8 Check settings 
and change 
them as 
needed.
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value: 
Hold value
Status: 
Bad and 
Constant
(0x30)
Illegal ET 
URV
Illegal ET 
SPAN
AL.57 
ET. ADJ
ET SPAN 
trim err
7 Adjust settings 
and change 
them as 
needed.
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Measured 
value
Status:
Poor 
Accuracy 
and Not 
Limited
(0x40)
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Measured 
value
Status:
Poor 
Accuracy 
and Not 
Limited
(0x40)
ET ZERO 
trim err
--- ET Fixed 
Mode
11 Under 
Temperature Fix 
Mode. PV is ET
Leave from 
Temperature 
Fix Mode.
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Hold value
Status:
Good and 
Constant
(0xF0)
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Hold value
Status:
Good and 
Constant
(0xF0)
AL.79 
OV.DISP
(None)
(None)
Displayed value 
exceeds limit.
Check settings 
and change 
them as 
needed.
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
AL.90 
SIM
Flow 
Simulation 
mode
5 Under Simulation 
Mode for ow.
Check 
Simulation 
Mode.
Value: Measured value
Status: Good and Not Limited
(0xC0)
or
Value: Simulation value
Status: Manual/Fixed and Constant
(0xB0)
Value: Measured value
Status: Good and Not Limited
(0xC0)
or
Value: Simulation value
Status: Manual/Fixed and 
Constant
(0xB0)
AL.91 
P. SIM
P Simulate 
Mode
11 Under Simulation 
Mode for device 
variables.
Check 
Simulation 
Mode
Value and
Status:
Pressure
simulation
value
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
unsupported
AL.91 
SP.SIM
SP 
Simulate 
Mode
Value and
Status:
Static
Pressure
simulation
value
AL.91 
ET.SIM
ET 
Simulate 
Mode
Value:
Measured value
Status:
Good and Not Limited
(0xC0)
Value and
Status:
Temperature
simulation
value