    <3.  Parameter Setting>
3-13
IM 01C25R05-01EN
3.4.6  Software Write Protection
EJX multivariable transmitter congured data is 
saved by using a write protection function. The 
write protection status is set to “Yes” when 8 
alphanumeric characters are entered in the New 
password eld and transferred to the transmitter.
When write protection is set to ”Yes,” the transmitter 
does not accept all the write possible parameters 
changes. When the same eight alphanumeric string 
entered in the New password eld is also entered 
in the Enable wrt 10min eld and transferred to the 
transmitter, it will be possible to change transmitter 
parameters during a 10 minute period.
To change the transmitter from the write protection 
”Yes” status back to write protection ”No” status, 
use Enable wrt 10min to rst release the write 
protection function and then enter eight spaces in 
the New password eld.
•  Procedure to call up the display 
[Root Menu] → Detailed setup → Device information 
→ Field device info → Wrt protect menu →
→ Write Protect Display current protect mode 
(Yes: protected, No: not protected)
→  Enable wrt 10 
min
Release the protect function for 10 
min.
→ New password Set the new password or change 
the password
3.4.7  Alarm
The function is used to display the alarm codes 
when the input differential pressure exceeds the 
specied value within the calibration range. The 
same is available for the input static pressure and 
external temperature. Refer to Table 4.1 Alarm 
Message Summary for the specic alarm code to 
be generated.
(1)  Alarm Setting
Select the process variable at Process Alert which 
the alarm is set, then set the alert mode for that 
value.
•  Procedure to call up the display
[Root Menu] → Detailed setup →Output condition → 
Process Alerts →
Selection of 
the process 
variable for 
alarm
→ Pres Alert Mode: Differential 
pressure
→ SP Alert Mode: Static pressure
→ ET Alert Mode: External 
temperature
Selection of  
alert mode 
Off: Disable the alert function
Hi Al Detect: High side alert detection
Lo Al Detect: Low side alert detection
Hi/Lo Al Detect: High and Low side 
alert detection
(2)  Threshold Level Setting
Set the threshold of high and low alert value for 
alarm generation.
•  Procedure to call up the display
[Root Menu] → Detailed setup →Output condition → 
Process Alerts →
Parameter Detail
→  Pres Hi Alert 
Val
Set the threshold value of upper side 
for differential pressure
→  Pres Lo Alert 
Val
Set the threshold value of lower side 
for differential pressure
→  SP Hi Alert 
Val
Set the threshold value of upper side 
for static pressure
→  SP Lo Alert 
Val
Set the threshold value of lower side 
for static pressure
→  ET Hi Alert 
Val
Set the threshold value of upper side 
for external temperature
→  ET Lo Alert 
Val
Set the threshold value of lower side 
for external temperature