User guide
    <3.  Parameter Setting>
3-12
IM 01C25R05-01EN
<3>  Enter the correction value of “2” to the Pres 
LTD (Manual, Lower Pt).
<4>  Call up the Pres UTD (Manual, Upper Pt).
<5>  Suppose that a standard pressure of 3000 
mmH
2
O is applied and the value of the Pres for 
trim is 3015.0. Firstly, obtain the slope error for 
the span as follows;
Slope Error =
Applied Pressure Value−Value of Pres for Trim
× (URV−LRV)
Applied Pressure Value
=
3000−3015
× (3000−1000) = −10
3000
Then correct for this slope error of −10 by adding 
−10 to Pres UTD (Manual, Upper Pt). 
  −3.0+(−10.0)=−13.0
<6>  Enter the correction value of “−13” to the Pres 
UTD (Manual, Upper Pt).
(3)  Sensor Trim for Static Pressure or External 
Temperature
For the EJX multivariable transmitter, full sensor 
trim of the static pressure or external temperature is 
performed in the same way as with the differential 
pressure.
•  Procedure to call up the display for static 
pressure
[Root Menu] → Diag/Service → Calibration → SP 
sensor trim → SP trim →
→ Auto, Lower Pt Auto trim for 0% point
→ Auto, Upper Pt Auto trim for 100% point
→ Manual, Lower Pt Manual trim for 0% point
→ Manual, Upper Pt Manual trim for 100% point
•  Procedure to call up the display for external 
temperature
[Root Menu] → Diag/Service → Calibration → ET 
Sensor trim → ET trim →
→ Auto, Lower Pt Auto trim for 0% point
→ Auto, Upper Pt Auto trim for 100% point
→ Manual, Lower Pt Manual trim for 0% point
→ Manual, Upper Pt Manual trim for 100% point
(4)  Reset Trim Adjistment to Factory Setting
The Clear P trim, Clear SP trim and Clear ET trim 
commands can reset the trim adjustment to the 
initial calibrated values that were set. The amount of 
the adjustment performed with the external zero-
adjustment screw is returned to the initial setting as 
well.
•  Procedure to call up the display for differential 
pressure
[Root Menu] → Diag/Service → Calibration → Pres 
Sensor trim → Clear P trim → Execute
•  Procedure to call up the display for static 
pressure 
[Root Menu] → Diag/Service → Calibration → SP 
Sensor trim → Clear SP trim → Execute
•  Procedure to call up the display for external 
temperature 
[Root Menu] → Diag/Service → Calibration → ET 
Sensor trim → Clear ET trim → Execute
(5)  Calibration ag
When the Calibration Flag parameter is set, 
the corresponding status bit in Status Bits and 
Status Bytes parameter in Address Map (Basic 
Information) will turn ON. The Calibration Flag 
parameter is used only for informational use and 
does not affect the internal operation of the EJX 
Multivariable transmitter. Regarding the Calibration 
Flag parameter, refer to 4.1.1.
•  Procedure to call up the display
[Root Menu] →Diag/Service → Calibration 
→Calibration Flag
→ Calibration Flag Flag the transmitter is in a 
calibration State.
3.4.5  External Switch Mode
Follow the procedure below to enable or inhibit zero 
point adjustment by means of the zero-adjustment 
screw on the transmitter.
This is set to “Disable” when the instrument is 
shipped.
To change the mode, follow the procedure below. 
•  Procedure to call up the display
[Root Menu] → Detailed setup → Device information 
→ Field device info → Ext SW
Enabled Enable the external zero point 
adjustment
Disabled Disable the external zero point 
adjustment